_________ is a redirection of US in many directions as a result of a *rough* boundary between
two media - RIGHT ANSWER -scattering
___________ resolution is determined by SPL - RIGHT ANSWER -radial or axial
______________ is the distance covered by one cycle - RIGHT ANSWER -wavelength
______________ is the time to complete one cycle - RIGHT ANSWER -period
A circular hyperechoic structure is located at a depth of 5 cm. The structure has abnormally low
attenuation and the speed in the mass is 1,400 m/s. Which of the following is most likely to
appear beneath the structure? - RIGHT ANSWER -reflectors beneath the structure appear
abnormally deep in the image
A Doppler exam is performed with a 5 MHz probe and a PRF of 15 kHz. Which of these Doppler
shifts will create aliasing? - RIGHT ANSWER -14 kHz
A hydrophone is used to measure: - RIGHT ANSWER -intensity (or frequency)
A pulse is emitted by a transducer and is traveling in soft tissue. The go-return-time, or
time-of-flight, of a sound pulse is 130 microseconds. What is the reflector depth? - RIGHT
ANSWER -10 cm
A quality assurance program: - RIGHT ANSWER -is needed by all labs
A shadow appears on an ultrasound scan. Which of the following statements is true about the
shadow artifact? - RIGHT ANSWER -it appears as hypoechoic
,A shadow appears on an ultrasound scan. Which of the following statements is true about the
structure immediately above the shadowing artifact? - RIGHT ANSWER -attenuation properties
are too high
A single element transducer is focused: - RIGHT ANSWER -by the manufacturer
A sound beam is created by a single, disc-shaped crystal. Which of these beams has the
shallowest focus? - RIGHT ANSWER -2 MHz, 9 mm diameter
A sound wave is traveling in the body and propagating from one form of soft tissue to another.
For example, sound is moving from muscle to blood. What percentage of the sound wave is most
likely to be reflected at the muscle-blood boundary? - RIGHT ANSWER -.5%
A sound wave with a frequency of 5 MHz is travelling in soft tissue. What is the wavelength? -
RIGHT ANSWER -.3 mm
A sound wave's frequency is 10 MHz. The wave is traveling in soft tissue. What is its wavelength?
- RIGHT ANSWER -.15 mm
A system which is not working well now does not display clear images in the far field. This may
be a problem with: - RIGHT ANSWER -transducer, receiver, monitor, or scan converter
A transducer converts __________ from one form to another - RIGHT ANSWER -energy
A transducer with more than one active element is called: - RIGHT ANSWER -array
According to AIUM, there have been no confirmed bioeffects below intensities of ______ watts
per square centimeter SPTA - RIGHT ANSWER -.01
,Aliasing is an artifact observed when performing Doppler ultrasonography. Which of the
following signals is most likely to alias? - RIGHT ANSWER -7 MHz with a PRF of 3 kHz
All of the following are limitations of analog scan converts EXCEPT: - RIGHT ANSWER -low spatial
resolution
All of the following are primarily used to change the overall brightness of an entire image EXCEPT:
- RIGHT ANSWER -compensation
All of the following are true EXCEPT: - RIGHT ANSWER -two waves with identical frequencies
must interfere constructively
All of the following are true EXCEPT: - RIGHT ANSWER -two waves with identical frequencies
must interfere constructively
All of the following are true of normal cavitation EXCEPT: - RIGHT ANSWER -harmful bioeffects
are commonly observed
All of the following are true regarding temporal resolution EXCEPT: - RIGHT ANSWER -decreased
temporal resolution is associated with shorter pulse duration
All of the following artifacts place the erroneous reflection below the true reflector EXCEPT: -
RIGHT ANSWER -side lobe
All of the following describe dynamic range EXCEPT: logarithmic, comparison, relative, absolute,
ratio - RIGHT ANSWER -absolute
, All of the following regarding the hemodynamics of a stenosis are true EXCEPT: - RIGHT ANSWER
-pressure is lowest downstream
All of the following result from the use of backing material in a transducer EXCEPT: - RIGHT
ANSWER -high sensitivity
All of the following statements regarding in-vitro bioeffects are true EXCEPT: - RIGHT ANSWER
-in-vitro bioeffects conclusions are clinically relevant
All of the following will improve temporal resolution EXCEPT: - RIGHT ANSWER -increased line
density
An image is too dark. Which technique is best to correct this problem? - RIGHT ANSWER
-increase receiver gain
An individual is standing. The blood pressure generated by the heart is 120 mmHg. What is the
hydrostatic pressure at the ankle? - RIGHT ANSWER -100 mmHg
An ultrasound system is imaging to a depth of 10 cm. Which of the following will be different on
a single image at depths of 2 cm, 5 cm, and 10 cm? - RIGHT ANSWER -lateral resolution
An ultrasound system is operating in the harmonic mode. The image on the display of the
ultrasound system was created using sound reflections with a frequency of 6 MHz. What is the
frequency of sound emitted by the system's transducer? - RIGHT ANSWER -3 MHz
An ultrasound system is set at 0 dB and is transmitting at full intensity. What is the output power
when the system is transmitting at 10% of full intensity? - RIGHT ANSWER --10 dB
An ultrasound system is set at 0 dB and is transmitting at full intensity. What is the output power
when the system is transmitting at 50% of full intensity? - RIGHT ANSWER --3 dB