Questions and Answers 100% Pass
B. Authority having jurisdiction - ✔✔An existing system that is not in strict compliance
with the provisions of this code shall be permitted to be continued in a use as long as
the __________ has determined that such use does not constitute a distinct hazard to
life.
A. Architect
B. Authority having jurisdiction
C. Engineer
D. Nursing
B. Designated storage area - ✔✔Indoor locations for oxygen, nitrous oxide, and
mixtures of these gases shall not communicate with any of the following
except__________
A. Areas with open flames
B. Designated storage area
Brittie Donald, All Rights Reserved © 2025 1
,C. Engines
D. Kitchens
D. Visual pressure indicator - ✔✔An auxiliary source connection shall consist of all the
following except
A. Removable plug or cap
B. Tree
C. Valve
D. Visual pressure indicator
A. Each building served - ✔✔When multiple buildings are served from a single
cryogenic liquid oxygen source, it is required to place the emergency oxygen supply
connection in-
A. Each building served
B. The building closest to the source
C. The largest building served
D. The main building only
B. 3 - ✔✔The minimum clearance around the EOSC for a temporary source of supply is
____________ ft.
Brittie Donald, All Rights Reserved © 2025 2
, A. 2
B. 3
C. 5
D. 6
A. Except be filtered to 0.01 microns - ✔✔The quality medical air shall comply with all
of the following standards except
A. Be filtered to 0.01 microns
B. Have lesson than 25 ppm gaseous hydrocarbons
C. Have no detectable liquid hydrocarbons
D. Meet the requirement of medical air USP
B. Human respiratory applications - ✔✔Medical air sources shall only be use for _____
A. Air operating devices
B. Human respiratory applications
C. Instrument air
D. WAGD
Brittie Donald, All Rights Reserved © 2025 3