What is the Area Temp & Humidity for Decontamination? -
✔✔ANSW✔✔..60 deg to 65 deg F, 30% to 60% Humidity
What is the air exchange rate & pressure type for
Decontamination? - ✔✔ANSW✔✔..10 air exchanges per
hour under Negative Pressure
What is the Area Temp & Humidity for the Sterile Storage
Area? - ✔✔ANSW✔✔..68 deg to 75 deg F, 35% to 70%
Humidity
What is the air exchange rate & pressure type for Sterile
Storage Area? - ✔✔ANSW✔✔..4 air exchanges per hour
under Positive Pressure
What is the Area Temp & Humidity for the Prep & Packaging
Area? - ✔✔ANSW✔✔..68 deg to 73 deg F, Ideal Humidity
50% - not lower than 35%
What is the air exchange rate & pressure type for the Prep &
Packaging Area? - ✔✔ANSW✔✔..10 air exchanges per hour
under Positive Pressure
,The main role of sterile processing is to: -
✔✔ANSW✔✔..Provide dependable reliable services to
enhance the quality of patient care.
Instrument sets and patient care equipment must be
processed and distributed in an accurate and timely manner
so that: - ✔✔ANSW✔✔..Patient care is not adversely
affected.
Ethics is the practice of: - ✔✔ANSW✔✔..Doing the right
thing.
Compliance with OSHA is: - ✔✔ANSW✔✔..Required by law.
Which of the following establishes sterilization practice
standards based on opinion, research, and/or theory: -
✔✔ANSW✔✔..AAMI
Recommended practices are: - ✔✔ANSW✔✔..Statements of
sound principals of practice that are based on scientific
data and the opinions of experts.
EPA, FDA, OSHA and state departments of health (DOH)
primarily issue: - ✔✔ANSW✔✔..Regulations.
Standards for CS/SPD practice are written by: -
✔✔ANSW✔✔..AAMI
The Health Insurance Portability and Accountability Act
provides for which ethical issue? -
✔✔ANSW✔✔..Confidentiality
,A rule designed to govern behavior is called a: -
✔✔ANSW✔✔..Regulation
An established norm which is based on theory is called a: -
✔✔ANSW✔✔..Standard
Before a pesticide can be marketed and used in the United
States, it must be evaluated and registered by which of the
following? - ✔✔ANSW✔✔..EPA
Which of the following regulates and makes sure products
are safe and effective before companies put them on the
market? - ✔✔ANSW✔✔..FDA
Which sterilization method is suitable for the following
packaging materials: textiles, non-wovens, polyolefin wraps,
paper-plastic pouches and rolls, and rigid sterilization
containers - ✔✔ANSW✔✔..Steam
Which sterilization method is suitable for the following
packaging materials: textiles, non-wovens, polyolefin wraps,
paper-plastic pouches and rolls, Tyvek (all-plastic) pouches,
polyethylene, and most rigid sterilization containers -
✔✔ANSW✔✔..Ethylene Oxide
Which sterilization method is suitable for the following
packaging materials: polyolefin wraps, Tyvek (all-plastic)
pouches, and certain rigid sterilization containers -
✔✔ANSW✔✔..Low-Temperature Gas Plasma (LTGP)
Which sterilization method is suitable for the following
packaging materials: Metal and glass containers, aluminum
foil, glassine envelopes - ✔✔ANSW✔✔..Dry Heat
, Which sterilization method is suitable for the following
packaging materials: Non-woven cellulosic wrap (paper)
and certain containers - ✔✔ANSW✔✔..Ozone
Which sterilization method is suitable for the following
packaging materials: polyolefin wraps and Tyvek (all-
plastic) pouches that have been validated for use with low-
temperature hydrogen peroxide - ✔✔ANSW✔✔..Low-
Temperature Hydrogen Peroxide (V-Pro 1 Low Temperature
Sterilization System)
The unit calls for a patient-cooling machine. Which of the
following would be sent? - ✔✔ANSW✔✔..Hypo/Hyperthermia
machine
Which of the following pieces of equipment is used in the
prevention of blood clots? - ✔✔ANSW✔✔..Sequential
compression device
Patient care equipment MUST be processed: -
✔✔ANSW✔✔..After each use
Before being disinfected, patient care equipment should be:
- ✔✔ANSW✔✔..Cleaned with detergent
What is the BEST way to transport an IV Pump to CS/SPD
after patient use? - ✔✔ANSW✔✔..In a covered cart or
closed bag
Which disinfectant is to be used for patient care equipment?
- ✔✔ANSW✔✔..The disinfectant recommended by the
device manufacturer
verified answers/ complete solutions|get it 100%
accurate
What is the Area Temp & Humidity for Decontamination? -
✔✔ANSW✔✔..60 deg to 65 deg F, 30% to 60% Humidity
What is the air exchange rate & pressure type for
Decontamination? - ✔✔ANSW✔✔..10 air exchanges per
hour under Negative Pressure
What is the Area Temp & Humidity for the Sterile Storage
Area? - ✔✔ANSW✔✔..68 deg to 75 deg F, 35% to 70%
Humidity
What is the air exchange rate & pressure type for Sterile
Storage Area? - ✔✔ANSW✔✔..4 air exchanges per hour
under Positive Pressure
What is the Area Temp & Humidity for the Prep & Packaging
Area? - ✔✔ANSW✔✔..68 deg to 73 deg F, Ideal Humidity
50% - not lower than 35%
What is the air exchange rate & pressure type for the Prep &
Packaging Area? - ✔✔ANSW✔✔..10 air exchanges per hour
under Positive Pressure
,The main role of sterile processing is to: -
✔✔ANSW✔✔..Provide dependable reliable services to
enhance the quality of patient care.
Instrument sets and patient care equipment must be
processed and distributed in an accurate and timely manner
so that: - ✔✔ANSW✔✔..Patient care is not adversely
affected.
Ethics is the practice of: - ✔✔ANSW✔✔..Doing the right
thing.
Compliance with OSHA is: - ✔✔ANSW✔✔..Required by law.
Which of the following establishes sterilization practice
standards based on opinion, research, and/or theory: -
✔✔ANSW✔✔..AAMI
Recommended practices are: - ✔✔ANSW✔✔..Statements of
sound principals of practice that are based on scientific
data and the opinions of experts.
EPA, FDA, OSHA and state departments of health (DOH)
primarily issue: - ✔✔ANSW✔✔..Regulations.
Standards for CS/SPD practice are written by: -
✔✔ANSW✔✔..AAMI
The Health Insurance Portability and Accountability Act
provides for which ethical issue? -
✔✔ANSW✔✔..Confidentiality
,A rule designed to govern behavior is called a: -
✔✔ANSW✔✔..Regulation
An established norm which is based on theory is called a: -
✔✔ANSW✔✔..Standard
Before a pesticide can be marketed and used in the United
States, it must be evaluated and registered by which of the
following? - ✔✔ANSW✔✔..EPA
Which of the following regulates and makes sure products
are safe and effective before companies put them on the
market? - ✔✔ANSW✔✔..FDA
Which sterilization method is suitable for the following
packaging materials: textiles, non-wovens, polyolefin wraps,
paper-plastic pouches and rolls, and rigid sterilization
containers - ✔✔ANSW✔✔..Steam
Which sterilization method is suitable for the following
packaging materials: textiles, non-wovens, polyolefin wraps,
paper-plastic pouches and rolls, Tyvek (all-plastic) pouches,
polyethylene, and most rigid sterilization containers -
✔✔ANSW✔✔..Ethylene Oxide
Which sterilization method is suitable for the following
packaging materials: polyolefin wraps, Tyvek (all-plastic)
pouches, and certain rigid sterilization containers -
✔✔ANSW✔✔..Low-Temperature Gas Plasma (LTGP)
Which sterilization method is suitable for the following
packaging materials: Metal and glass containers, aluminum
foil, glassine envelopes - ✔✔ANSW✔✔..Dry Heat
, Which sterilization method is suitable for the following
packaging materials: Non-woven cellulosic wrap (paper)
and certain containers - ✔✔ANSW✔✔..Ozone
Which sterilization method is suitable for the following
packaging materials: polyolefin wraps and Tyvek (all-
plastic) pouches that have been validated for use with low-
temperature hydrogen peroxide - ✔✔ANSW✔✔..Low-
Temperature Hydrogen Peroxide (V-Pro 1 Low Temperature
Sterilization System)
The unit calls for a patient-cooling machine. Which of the
following would be sent? - ✔✔ANSW✔✔..Hypo/Hyperthermia
machine
Which of the following pieces of equipment is used in the
prevention of blood clots? - ✔✔ANSW✔✔..Sequential
compression device
Patient care equipment MUST be processed: -
✔✔ANSW✔✔..After each use
Before being disinfected, patient care equipment should be:
- ✔✔ANSW✔✔..Cleaned with detergent
What is the BEST way to transport an IV Pump to CS/SPD
after patient use? - ✔✔ANSW✔✔..In a covered cart or
closed bag
Which disinfectant is to be used for patient care equipment?
- ✔✔ANSW✔✔..The disinfectant recommended by the
device manufacturer