SCORM Nova Stat Strip Training Video
If data transfer was incomplete before the meter was removed from the docking station,
it is still OK to use the meter - answer True
The purpose of the docking station is - answer Charge the battery and meter
Charge the spare battery
Download results
Upload new operator and patient information
What is the use life of the test strips after they have been opened? - answer180 days or
the printed expiration date on the vial, whichever comes first
What is the expiration date for the quality control material once opened - answer90 days
or the printed expiration date on the vial, whichever comes first
The sample must be applied to the end of the strip and must fill the strip completely
upon touching the quality control drop. If it does not fill completely, you may touch the
strip to the quality control drop a second time until the countdown begins - answer False
The meter should always be held level when applying the sample to the strip to keep
liquid from entering into the meter and damaging it - answerTrue
A padlock symbol next to the Glu box indicates - answerThe meter is locked out
because QC is due; perform QC testing or perform patient testing on an alternate meter
The meter requires QC to be run before allowing patient samples
Critical results are indicated by - answerResults highlighted by two up or down arrows
Batteries have a finite useful life and should be periodically checked for cracks or
swelling - answerTrue
The meter should be cleaned and disinfected - answerWhen visibly soiled
After use on an isolation patient
After every patient
If data transfer was incomplete before the meter was removed from the docking station,
it is still OK to use the meter - answer True
The purpose of the docking station is - answer Charge the battery and meter
Charge the spare battery
Download results
Upload new operator and patient information
What is the use life of the test strips after they have been opened? - answer180 days or
the printed expiration date on the vial, whichever comes first
What is the expiration date for the quality control material once opened - answer90 days
or the printed expiration date on the vial, whichever comes first
The sample must be applied to the end of the strip and must fill the strip completely
upon touching the quality control drop. If it does not fill completely, you may touch the
strip to the quality control drop a second time until the countdown begins - answer False
The meter should always be held level when applying the sample to the strip to keep
liquid from entering into the meter and damaging it - answerTrue
A padlock symbol next to the Glu box indicates - answerThe meter is locked out
because QC is due; perform QC testing or perform patient testing on an alternate meter
The meter requires QC to be run before allowing patient samples
Critical results are indicated by - answerResults highlighted by two up or down arrows
Batteries have a finite useful life and should be periodically checked for cracks or
swelling - answerTrue
The meter should be cleaned and disinfected - answerWhen visibly soiled
After use on an isolation patient
After every patient