ACNS Guidelines Questions with Verified Correct
Answer
Monitoring physician

Should be licensed in the state and privileged to interpret neurophysiologic testing in the

hospital. They are responsible for interpretation of data.

How many cases is it recommended the online physician monitor at once?

No more than three cases concurrently

What should monitoring system include?

the capacity to:

display raw data and continuous trends

store time/date of recorded waveforms

trigger stim of potentials such as EMG, peripheral or cranial nerve

record free run signals

record technologist's notes with waveform data

retrieve and review recorded data for post-hoc analysis

automatic artifact rejection

remote, real time display of data

Can post-hoc analysis replace real time interpretation by physician?

No

What is the limit on total chassis leakage?

,100 uA

What is the limit on chassis leakage for isolated patient connections?

10 uA

The grounding system should ensure that no voltage greater than ________ can be

measured across an impedance of ______________, typically at _____ Hz, between any

combo of exposed or conductive surfaces or grounding conductors that may be in contact

with patient or personnel.

20 mV rms, 1000 ohms, 60 Hz

How often must IONM systems be inspected?

At least twice a year

True or false: interpreter should be cognizant that changes in data may be just as marked

with hypotension and cardiovascular collapse as with surgical manipulation or

compression of neural pathways

True

True or false: N20 and halogenated inhalational agents can have even more prominent

attenuation effects on lower extremity SSEPs and in young children

True

When are baseline IONM values established?

,Following induction of anesthesia and positioning of patient, and prior to surgical intervention.

Baseline values may need to be reestablished if changes in anesthetic medications or other

physiological parameters occur during the case.

True or false: In cases where it is determined that there is little or no risk to neural

structures during a specific portion of a particular procedure, the interpreting physician

may, in consulation with the surgeon , elect to suspend IONM

true

True or false: In general, surgical injury more often produced waveform amplitude or

morphology changes, than latency prolongation.

True

True or false: Gain should be adjustable in steps of not more than 2.5 to 1

True

Differential input impedance of amplifier must be at least ___________

100 megaOhm

Common mode rejection should be _________

at least 80 dB (10,000:1)

True or false: With filters wide open, the amplifier bandpass measured at the -3 dB points

must be at least 0.1-5,000 Hz. The rolloff slops must be specified.

True

The noise level of the amplifier must not exceed _______

, 2 uV rms with all the inputs connected to neutral and with a bandpass of 0.1-5,000 Hz

Time (horizontal) resolution of the EP system should be ____ us/data point of less

20 us/data point

Amplitude resolution of eight bits or 12 bits at the A-D converted?

eight bits adequate for many applications, but 12 bit converters are preferred because they are

usable across a broader dynamic range

How many addresses of memory should be available for each channel?

500

How many trials should the evoked potential system allow for?

at least 4,000

True or false: excessive analog filtering will alter the latency, amplitude, and morphology of

evoked potential components.

True

True or false: Excessive low-frequency filtering will cause a peak or valley to appear

earlier, leading to falsely normal results.

True (Phase lead)

True or false: Excessive high-frequency filtering will delay (prolong) a peak or valley

leading to a possible falsely abnormal study

True: phase lag

Document information

Uploaded on
October 29, 2026
Number of pages
35
Written in
2026/2027
Type
Exam (elaborations)
Contains
Questions & answers
$15.99

Wrong document? Swap it for free Within 14 days of purchase and before downloading, you can choose a different document. You can simply spend the amount again.
Written by students who passed
Immediately available after payment
Read online or as PDF

Seller avatar
Reputation scores are based on the amount of documents a seller has sold for a fee and the reviews they have received for those documents. There are three levels: Bronze, Silver and Gold. The better the reputation, the more your can rely on the quality of the sellers work.
KenAli
2.6
(18)
Sold
121
Followers
5
Items
25960
Last sold
2 days ago



Why students choose Stuvia

Created by fellow students, verified by reviews

Quality you can trust: written by students who passed their tests and reviewed by others who've used these notes.

Didn't get what you expected? Choose another document

No worries! You can instantly pick a different document that better fits what you're looking for.

Pay as you like, start learning right away

No subscription, no commitments. Pay the way you're used to via credit card and download your PDF document instantly.

Student with book image

“Bought, downloaded, and aced it. It really can be that simple.”

Alisha Student

Working on your references?

Create accurate citations in APA, MLA and Harvard with our free citation generator.

Working on your references?

Frequently asked questions