and Answers (Verified Answers) Most Recent
exam COMPLETE (2026) (Latest Update 2026)
UPDATE!!
Which of the following is varied automatically when using automatic
exposure control (AEC)?
ANSWERS: Exposure time
Which of the following can cause too much heat on the anode and
cause it to crack and fail?
1. excessive exposure on a cold tube
2. a rapid series of large exposures
3. mA settings greater than 400 mA
ANSWERS: 1 and 2
X-ray tubes should undergo the warm-up procedure again after
they have been idle for more than:
ANSWERS: 1 hour
,What is the heat unit (HU) of the anode if 200 mA, 0.10 sec., and 80
kVp is used on a single- phase generator?
ANSWERS: 1600
How many heat units (HU) are generated on a high frequency (HF)
generator if an exposure time of 300 mA, 0.12 sec., and 95 kVp is
used?
ANSWERS: 3,420
Which of the following will help prolong the life of an x-ray tube?
1. warm up the anode each morning
2. use low ma setting whenever possible
3. use the low-speed rotor whenever possible
ANSWERS: 1,2, and 3
Which of the following x-ray generators produces the greatest
amount of x-rays for the same exposure technique?
ANSWERS: High-frequency
Radiographic contrast is controlled by the:
ANSWERS: kVp
The quantity, or number of photons produced per second during an
exposure is controlled by the:
ANSWERS: mA
,The variation in tissue densities within the patient is referred to as
the:
ANSWERS: Subject contrast
The overall contrast seen in the radiographic image is referred to as:
ANSWERS: Radiographic contrast
Which of the following are considered a part of shape distortion?
1. magnification
2. foreshortening
3. elongation
ANSWERS: 2 and 3
The unsharp edges of a body part are referred to as the:
ANSWERS: Penumbra
Which of the following would be considered involuntary motion?
1. tremors
2. peristalsis
3. heartbeats
ANSWERS: 1, 2, and 3
Quantum mottle is typically seen on the radiographic image when
using:
ANSWERS: high mA, high kVp techniques
, The primary controlling factor of penetration of x-rays is:
ANSWERS: kVp
The term used in the new digital imaging environment to replace
density is:
ANSWERS: Brightness
The geometric factors that affect the formation of the image are:
1. SID
2. OID
3. Focal spot
ANSWERS: 1, 2, and 3
When using computed radiography (CR), where is the latent image
stored until it is processed?
ANSWERS: In a photostimulable phosphor plate
Digital imaging requires that the exposure technique is accurately
set. Which of the following is the most critical of the exposure
factors?
ANSWERS: kVp
One of the major advantages of using CR or DR systems is the:
ANSWERS: The ability to see images very fast