Question:
NFPA 99 MEDGAS FOR 6010 INSTALLER
25 ppm
Answer:
The ppm gaseous hydrocarbons in the medical air
system shall be less than?
Question:
32 degrees F
Answer:
At any level of demand, the medical air dryers
shall provide air with a dew point below this temperature?
Question:
medical air compressors
Answer:
Which of the following is NOT
permitted to have the system supply source located outside?
Question:
A. vacuum pumps
B. manifolds for gas cylinders
C. instrument air compressors
D. medical air compressors
duplexed
Answer:
Regulators, filters, and dryers shall be sized to meet
the peak calculated demand. The piping format for these items shall be at a minimum?
,Question:
Authority Having Jurisdiction (AHJ)
Answer:
They can allow an existing
system, which is not in compliance with the NFPA 99 Code to remain in service.
Question:
pressure regulator
Answer:
Medical air supply systems operating at
different pressures shall be provided with?
Question:
unusable
Answer:
The removal of the outlet connection on a cryogenic
liquid container will cause the container to become?
Question:
laboratory
Answer:
A building, space, room, or group of rooms intended
to serve activities involving procedures for investigation, diagnosis, or treatment in which
flammable, combustible, or oxidizing materials are to be used.
Question:
20 degrees F
Answer:
Carbon dioxide and nitrous oxide cylinders shall be
kept from reaching temperatures below .
Question:
-22 degrees F
Answer:
The master alarm shall activate when the
instrument air compressor dew point temperature exceeds .
,Question:
Category 1
Answer:
In this space, alarm sensors shall be installed on the
patient's side of each individual zone valve box.
Question:
patient care vicinity
Answer:
The locations within 6' of the patient's bed
and 7'-6" from the ceiling, where examinations take place?
Question:
high water level
Answer:
Where liquid ring compressors for medical air
are used, the air receiver shall have a local alarm which will shut down the system due to?
Question:
Category 4 Space
Answer:
An area where failure of a system is not likely
to cause any physical impact on a patient.
Question:
hyperbaric facility
Answer:
Building or structure, used to house
hyperbaric chambers for medical applications and procedures at pressures above normal
atmospheric pressure.
Question:
area alarm
Answer:
The warning system that provides a continuous
, visible and audible vigilant watch on Category 1 & Category 2 piping systems.
Question:
500
Answer:
Moisture concentration of the outlet testing during the
verifier purity test shall not exceed how many ppm?
Question:
local alarm
Answer:
This alarm shall be triggered once the backup
instrument air compressor is energized.
Question:
1000 degrees F
Answer:
The components which make up surface
mounted medical gas rails shall have a melting point of no less than .
Question:
10
Answer:
When the carbon monoxide level in a medical air system
exceeds this ppm level, the local alarm is triggered.
Question:
5 ppm
Answer:
The most remote outlet & the source gas undergo the
verifiers piping purity test. The ppm non-methane variation between the two shall not exceed?
Question:
hospital
Answer:
Building used on a 24-hour basis for medical,
Questions and Answers | Latest Updated Study Material 2026
Question:
NFPA 99 MEDGAS FOR 6010 INSTALLER
25 ppm
Answer:
The ppm gaseous hydrocarbons in the medical air
system shall be less than?
Question:
32 degrees F
Answer:
At any level of demand, the medical air dryers
shall provide air with a dew point below this temperature?
Question:
medical air compressors
Answer:
Which of the following is NOT
permitted to have the system supply source located outside?
Question:
A. vacuum pumps
B. manifolds for gas cylinders
C. instrument air compressors
D. medical air compressors
duplexed
Answer:
Regulators, filters, and dryers shall be sized to meet
the peak calculated demand. The piping format for these items shall be at a minimum?
,Question:
Authority Having Jurisdiction (AHJ)
Answer:
They can allow an existing
system, which is not in compliance with the NFPA 99 Code to remain in service.
Question:
pressure regulator
Answer:
Medical air supply systems operating at
different pressures shall be provided with?
Question:
unusable
Answer:
The removal of the outlet connection on a cryogenic
liquid container will cause the container to become?
Question:
laboratory
Answer:
A building, space, room, or group of rooms intended
to serve activities involving procedures for investigation, diagnosis, or treatment in which
flammable, combustible, or oxidizing materials are to be used.
Question:
20 degrees F
Answer:
Carbon dioxide and nitrous oxide cylinders shall be
kept from reaching temperatures below .
Question:
-22 degrees F
Answer:
The master alarm shall activate when the
instrument air compressor dew point temperature exceeds .
,Question:
Category 1
Answer:
In this space, alarm sensors shall be installed on the
patient's side of each individual zone valve box.
Question:
patient care vicinity
Answer:
The locations within 6' of the patient's bed
and 7'-6" from the ceiling, where examinations take place?
Question:
high water level
Answer:
Where liquid ring compressors for medical air
are used, the air receiver shall have a local alarm which will shut down the system due to?
Question:
Category 4 Space
Answer:
An area where failure of a system is not likely
to cause any physical impact on a patient.
Question:
hyperbaric facility
Answer:
Building or structure, used to house
hyperbaric chambers for medical applications and procedures at pressures above normal
atmospheric pressure.
Question:
area alarm
Answer:
The warning system that provides a continuous
, visible and audible vigilant watch on Category 1 & Category 2 piping systems.
Question:
500
Answer:
Moisture concentration of the outlet testing during the
verifier purity test shall not exceed how many ppm?
Question:
local alarm
Answer:
This alarm shall be triggered once the backup
instrument air compressor is energized.
Question:
1000 degrees F
Answer:
The components which make up surface
mounted medical gas rails shall have a melting point of no less than .
Question:
10
Answer:
When the carbon monoxide level in a medical air system
exceeds this ppm level, the local alarm is triggered.
Question:
5 ppm
Answer:
The most remote outlet & the source gas undergo the
verifiers piping purity test. The ppm non-methane variation between the two shall not exceed?
Question:
hospital
Answer:
Building used on a 24-hour basis for medical,