SECTION 1: CLEANING, DECONTAMINATION & DISINFECTION Q1 – Q10
══════════════════════════════════════
Question 1 of 50
A 45-year-old CST is manually cleaning a set of orthopedic instruments in the
decontamination area. One of the chisels has dried blood and bone debris in the
serrations. She has already sprayed it with enzymatic foam at the point of use. Before
placing it in the ultrasonic cleaner, she should:
A. Soak it in plain water for 30 minutes to loosen debris
B. Brush it under running water with a soft-bristle brush to remove visible soil ✓
CORRECT
C. Place it directly in the ultrasonic cleaner because the enzymes will handle all soil
D. Wipe it with alcohol prep pads to dissolve the organic material
Correct Answer: B
Rationale: Visible soil must be removed manually before ultrasonic cleaning because
the ultrasonic bath is designed for fine cleaning of hidden lumens and joints, not for
,gross debris removal. Placing heavily soiled instruments directly into the ultrasonic can
damage the unit and redeposit soil onto other items. Pre-cleaning at the sink with a
brush is standard practice in every accredited sterile processing department.
Question 2 of 50
A decontamination technician in a 300-bed hospital notices that the temperature in the
soiled utility room where instruments await transport is 85°F. The humidity reading is
75%. According to AAMI standards, the most immediate concern with these
environmental conditions is that:
A. The enzymatic detergent will lose potency within two hours
B. The ultrasonic cleaner will require recalibration
C. The sterile storage area must match these conditions
D. High heat and humidity accelerate microbial growth and biofilm formation on soiled
instruments ✓ CORRECT
Correct Answer: D
Rationale: AAMI recommends keeping soiled instrument holding areas cool and dry to
prevent microbial proliferation and biofilm development, which make cleaning
significantly more difficult. While enzymatic detergents have stability concerns,
temperature and humidity in the holding area primarily affect microbial growth, not
chemical potency. Biofilm formation is one of the most serious challenges in central
service because it protects organisms and resists standard cleaning.
Question 3 of 50
A 28-year-old technician is preparing to decontaminate flexible endoscopes that arrived
from the endoscopy suite. The manufacturer's instructions specify high-level
disinfection. Before proceeding, he must verify that the disinfectant:
,A. Has been FDA-cleared for the specific endoscope model and contact time ✓
CORRECT
B. Is at room temperature regardless of the manufacturer's recommendation
C. Can be reused indefinitely as long as the solution remains clear
D. Requires only a 30-second immersion for high-level disinfection
Correct Answer: A
Rationale: High-level disinfectants used on flexible endoscopes must be FDA-cleared for
the specific device and used according to the labeled contact time and concentration;
using an unapproved agent can damage the scope and fail to kill mycobacteria and
spores. Endoscope high-level disinfection typically requires 5 to 20 minutes of contact
time, not seconds, and solutions have validated reuse periods with mandatory testing.
Technicians must always cross-reference the scope IFU with the disinfectant label.
Question 4 of 50
During a routine quality audit, a supervisor finds that a technician has been adding
enzymatic detergent to the ultrasonic bath at twice the manufacturer's recommended
concentration. The supervisor explains that over-concentration is problematic because:
A. It voids the ultrasonic cleaner's warranty automatically
B. It causes instruments to corrode immediately upon contact
C. It can leave a residue on instruments that interferes with steam sterilization ✓
CORRECT
D. It makes the solution too viscous to drain from the tank
Correct Answer: C
Rationale: Excess enzymatic detergent can leave a chemical film on instruments that
may interfere with steam penetration and contact during sterilization, potentially leading
to wet packs or sterilization failure. While corrosion is a concern with some chemicals,
enzymatic over-concentration primarily creates a residue problem. Residue also
increases the risk of patient reactions and can cause staining on surgical instruments.
, Question 5 of 50
A sterile processing technician is cleaning a laparoscopic instrument with a 5mm
diameter lumen. After flushing with enzymatic solution, she attaches a syringe to flush
the lumen with distilled water. The primary reason for using distilled or deionized water
for the final rinse is:
A. To neutralize the pH of the enzymatic detergent
B. To prevent mineral deposits and spotting that can occur with tap water ✓ CORRECT
C. To kill any remaining vegetative bacteria in the lumen
D. To reduce the surface tension of the water for faster drying
Correct Answer: B
Rationale: Tap water contains minerals that can leave deposits inside lumens and on
instrument surfaces, potentially interfering with function and sterilization. Distilled or
deionized water eliminates this mineral residue. The final rinse water is not primarily for
disinfection; that occurs later in the process. Mineral deposits in laparoscopic lumens
can obstruct channels and compromise instrument performance during surgery.
Question 6 of 50
A hospital's water treatment system fails, and the decontamination area must
temporarily use tap water for instrument rinsing. The director of sterile processing
knows that according to AAMI ST79, the greatest risk of using untreated tap water in
the final rinse is:
A. It will dissolve the lubricant applied after cleaning
B. It causes immediate rusting of all stainless steel instruments
C. It inactivates chemical indicators used in sterilization
D. It introduces endotoxins and pyrogens that can cause patient reactions ✓ CORRECT
Correct Answer: D