Medical Gas 6010 Installers Practice Test
Questions and Answers| Updated

An existing system that is not in strict compliance with the provisions of this code shall be
permitted to be continued in use as long as the _ has determined that such use does not
constitute a distinct hazard to life.

A) architect B) authority having jurisdiction C) engineer D) nursing staff

B) authority having jurisdiction




Cylinder and cryogenic liquid container outlet connections shall be designed as to render the
attachment point _ when removal of the outlet connection is attempted.

A) reusable B) unusable C) unsuitable D) universal

B) unusable




Cylinders in use and in storage shall be prevented from reaching temperatures in excess of _ *F.
A) 120 B) 125 C) 130 D) 135

B) 125




Nitrous oxide and carbon dioxide using cylinders or portable containers shall be prevented from
reaching temperatures lower than _ *F A) 130 B)125 C) 20 D) -20

C)20




Which of the following systems shall not be permitted to be located in an outdoor enclosure?

, A) bulk cryogenic liquid cylinders B) manifolds for gas cylinders C) manifolds for cryogenic liquid
containers D) medical air sources

D) medical air sources




Indoor locations for oxygen, nitrous oxide, and mixtures of these gases shall not communicate
with any of the following except _?

A) area with open flames B) designated storage area C) engines D) kitchens

B) designated storage area




Locations for central supply systems shall have _ devices protected from physical damage?

A) electrical B) pressure indicating C) regulating D) relief

A) electrical




Outdoor locations surrounded by impermeable walls, except fire barrier walls, shall have
protected ventilation openings at the _ of each wall to allow free circulation of air within the
enclosure?

A) base B) corner) C) middle D) top

A) base




All positive pressure supply systems shall be provided with duplex line pressure regulators piped
in _?

A) circuit B) loop C) parallel D) series

C) parallel

,All pressure relief valves shall be made of brass, bronze, or _?

A) carbon steel B) cast iron C) copper D) stainless steel

D) stainless steel




Central supply systems for positive pressure gases shall have one or more relief valves, the
valves shall have a relief setting that is _% above normal working operating pressure?

A) 30 B) 40 C) 50 D) 60

C) 50




When vented to the outside, relief valve vent lines shall be labeled in which way?

A) adhesive markers B) any manner that will distinguish them from the medical gas pipeline C)
stenciling D) stickers

B) any manner that will distinguish them from the medical gas pipeline




An auxiliary source connection shall consist of all of the following except?

A) removable plug or cap B) tee C) valve D) visual pressure indicator

D) visual pressure indicator




In central supply systems using gas or liquid cylinders, each header shall have a _ to prevent
debris from entering the manifold controls?

A) check valve B) filter C) monitor D) relief valve

, B) filter




Manifolds for gas cylinders shall consist of two equal headers, each with enough cylinders for an
average day's supply, but not fewer than _ connections?

A) 5 B) 4 C) 3 D) 2

D) 2




Manifolds for gas cylinders shall have a(n) _ signal that visibly indicates the operating status of
the equipment and shall activate an indicator at all master alarm panels when or just before
changeover from one header to the other takes place?

A) area B) master C) local D) zone

C) local




When multiple buildings are served from a single cryogenic liquid oxygen source, it is required
to place the emergency oxygen supply connection in _ ?

A) each building served B) the building closest to the source C) the largest building served D) the
main building only

A) each building served




The minimum clearance around EOSC for a temporary source of supply is _ ft?

A) 2 B) 3 C) 5 D) 6

B) 3

Document information

Uploaded on
May 21, 2026
Number of pages
36
Written in
2025/2026
Type
Exam (elaborations)
Contains
Questions & answers
$10.99

Wrong document? Swap it for free Within 14 days of purchase and before downloading, you can choose a different document. You can simply spend the amount again.
Written by students who passed
Immediately available after payment
Read online or as PDF

Seller avatar
Reputation scores are based on the amount of documents a seller has sold for a fee and the reviews they have received for those documents. There are three levels: Bronze, Silver and Gold. The better the reputation, the more your can rely on the quality of the sellers work.
KelvinBrooks
4.3
(94)
Sold
839
Followers
11
Items
5826
Last sold
1 day ago



Why students choose Stuvia

Created by fellow students, verified by reviews

Quality you can trust: written by students who passed their tests and reviewed by others who've used these notes.

Didn't get what you expected? Choose another document

No worries! You can instantly pick a different document that better fits what you're looking for.

Pay as you like, start learning right away

No subscription, no commitments. Pay the way you're used to via credit card and download your PDF document instantly.

Student with book image

“Bought, downloaded, and aced it. It really can be that simple.”

Alisha Student

Working on your references?

Create accurate citations in APA, MLA and Harvard with our free citation generator.

Working on your references?

Frequently asked questions