During a CT procedure, attenuation of the x-ray beam may be caused by which three
of the following? (Select three)
A. Coherent Scattering
B. Bremsstrahlung Radiation
C. Compton scattering
D. Photoelectric Absorption
E. Characteristic Radiation


Give this one a try later!


A. Coherent Scattering, C. Compton scattering, & D. Photoelectric
Absorption

During a CT procedure, attenuation of the x-ray beam may be caused by
coherent scattering, photoelectric absorption, and Compton scattering.
Attenuation refers to any interaction in which the x-ray beam decreases in
intensity. Coherent scattering, photoelectric absorption, and Compton
scattering all have the potential to remove photons from the useful beam
and cause attenuation. Bremsstrahlung and Characteristic radiation refer to

, x-ray production processes in the x-ray tube, not interaction mechanisms in
matter.




Concerning radiation-induced cancer, increasing the radiation dose to a patient
results in:
A. Increased severity of cancer
B. Decreased severity of cancer
C. Increase probability of cancer
D. Decreases probability of cancer


Give this one a try later!


C. Increase probability of cancer


Increasing radiation dose increases the probability of cancer, but it does
not increase the severity of cancer. A cancer caused by 3 Gray (Gy) of
radiation is no worse than a cancer caused by 2 Gy of radiation. Radiation
dose only affects the total probability.




Which of the following values represents a normal Computed Tomography Dose
Index (CTDI) of a pediatric CT head?
A. 75 milligray (mGy)
B. 85 milligray (mGy)
C. 40milligray (mGy)
D. 20 milligray (mGy)


Give this one a try later!

, C. 40milligray (mGy)

A normal value of the CTDI of a pediatric CT head is 40 mGy. According to
the American College of Radiology (ACR), a pediatric head scan on a 5
year old patient should not exceed 40 mGy. A CTDI higher than this level
may indicate that inappropriate technical parameters were used.




During a CT study, the technologist decreases the beam width (increases collimation).
What effect will this have on patient dose?
A. Increased patient dose
B. Decreased patient dose
C. No change in patient dose


Give this one a try later!


A. Increased patient dose


Decreasing the beam width actually increases the patient dose (this is
opposite of general radiography). This relates to the penumbra radiation
outside of the useful beam. When a smaller more collimated beam is used,
this creates more penumbra and an increase in patient dose.




While positioning a patient for CT of the chest, the technologist does not center the
patient in the center of the gantry. The scanner is using a variable milliamperage (mA).
This error is most likely to result in:
A. Increased patient dose
B. Decreased contrast resolution
C. Beam-hardening artifact
D. Increased image magnification


Give this one a try later!

, A. Increased patient dose

Incorrect patient positioning can cause an increase in patient dose when
using a variable mA. This is because the patient will look bigger on the
scout images and the scanner will increase the mA accordingly. There will
be no change in contrast resolution, no beam hardening, and no
magnification on the helical images.




A CT examination of the head used a fixed technique of 300 milliamperage (mA), 1.0
second rotation time, and 140 kilovoltage peak (kVp). The resulting Computed
Tomography Dose Index (CTDI) was 69 milligray (mGy). If the technique is changed to
120 kVp, how will the patient dose change?
A. Decrease patient dose
B. Increase patient dose
C. No change in patient dose


Give this one a try later!


A. Decrease patient dose

In this scenario, if the technique is changed to 120 kVp, the patient dose will
decrease. Changes in kVp have a significant effect on patient dose. As kVp
increases, the patient dose also increases.




How does decreasing kilovoltage (kVp) affect patient dose?
A. Increased patient dose
B. Decreased patient dose
C. No change in patient dose


Give this one a try later!

Document information

Uploaded on
February 23, 2026
Number of pages
35
Written in
2025/2026
Type
Exam (elaborations)
Contains
Questions & answers
$10.50

Wrong document? Swap it for free Within 14 days of purchase and before downloading, you can choose a different document. You can simply spend the amount again.
Written by students who passed
Immediately available after payment
Read online or as PDF

Seller avatar
EXAMROOM
5.0
(1)
Sold
9
Followers
0
Items
3095
Last sold
5 days ago




Why students choose Stuvia

Created by fellow students, verified by reviews

Quality you can trust: written by students who passed their tests and reviewed by others who've used these notes.

Didn't get what you expected? Choose another document

No worries! You can instantly pick a different document that better fits what you're looking for.

Pay as you like, start learning right away

No subscription, no commitments. Pay the way you're used to via credit card and download your PDF document instantly.

Student with book image

“Bought, downloaded, and aced it. It really can be that simple.”

Alisha Student

Working on your references?

Create accurate citations in APA, MLA and Harvard with our free citation generator.

Working on your references?

Frequently asked questions