SOLUTIONS GRADED A+
✔✔Peel pouches are an acceptable packaging material for ethylene oxide sterilization,
T/F? - ✔✔True
✔✔Name some rigid endoscopes - ✔✔Laraoscopes
Arthroscopes
Cystoscope
Sinuscope
Hysteroscope
Operating Laparoposcope
✔✔Name some semi-rigid endoscope - ✔✔Ureteroscope
✔✔Name some flexible endoscopes - ✔✔-Colonoscope
-Gastroscope
-Bronchoscope
-Duodenoscope
-Cystoscope
✔✔Endoscopes may be used for what purposes - ✔✔-Routine screening: Cancer
screening
-Diagnostic treatment: determine source of pain, bleeding, obtain biopsy specimen
-Therapeutic treatment: polyp, stone removal
✔✔What organization wrote the ST-91: Flexible and semi-rigid endoscope processing in
health care facilities. - ✔✔AAMI
✔✔All channels should be purged with filtered medical-grade air at the correct psi (as
per IFU), and that "before storage, the channel of the HLD endoscope should be dry to
help prevent bacterial growth and formation of biofilm. T/F - ✔✔True, According to
ANSI/AAMI ST-91.
✔✔According to AAMI ST91, what are two basic components of user verification of
cleaning efficacy. - ✔✔1. Establish a benchmark - a level of cleaning that can be
achieved consistently using specific soil markers.
2. Using rapid, easy-to-perform methods that reliably demonstrate that the cleaning
benchmarks have been achieved.
✔✔What are some cleaning verification test that can be used with flexible endoscopes.
- ✔✔-ATP tests
-Protien,hemoglobin and carbohydrate test.
, -Use of a Boroscope
✔✔Identify common errors associated with endoscope leak testing: - ✔✔1. Skipping
leak test to save time
2. Not following the endoscope IFU
3. Failure to prepare endoscope for immersion
4. Immersing endoscope in water that contain chemical solution
5. Not fully submerging the endoscope
6. Coiling the endoscope too tightly
7. Failure to fully angulate the distal tip
8. Failure to leak test for the required amount of time.
✔✔During point of use cleaning, technicians should flush the auxiliary water channel
and then detach the endoscope from the light source and suction pump. T/F - ✔✔True.
✔✔What are some common errors in point-of-use cleaning and handling? - ✔✔-Failure
to understand its importance.
-Lack of information
-Cleaning delays
-Lack of time
-Lack of communication
-Lack of proper equipment and supplies
✔✔Identify flexible endoscope sections - ✔✔-Light guide connector
-Light guide tube
-Video remote switches
-Control body
-Internal instrument channels
-Insertion tube
-Bending section/distal tip
✔✔Identify rigid endoscope parts - ✔✔a. Working length
b. length
c. distal end
d. field of view (FOV)
e. angle of view (AOV)
f. outside diameter
g. sheath mount
h. body
i. proximal end
j. light post adaptors
✔✔ANSI/AAMI ST91:2021 is an amendment to ST91:2015, T/F? - ✔✔False. it replaces
the older version.