EKG monitored patient should be __________, monitor chassis should be
___________ (nonconductive). Why - Answer - c ungrounded, grounded.
Grounded patient would be able to conduct leakage currents. we dont want this
lb to kg - Answer - c 2.2 lb = 1 kg F to C - Answer - c C=5/9(F-32) C to F -
Answer - c F=9/5C+32
3 things that change automaticity of automatic cells - Answer - c 1.
slope of phase 4
2. Change of threshold
potential
3. Change of resting membrane
potential
a systole is represented by what portion of the ECG - Answer - c QT interval
What is main ion transfer during systole - Answer - c K+ seeps out, Ca2+ enters
At rest, what is intracellular/extracellular environment - Answer - c inside
negative, outside positive
What phase of AP is resting membrane potential - Answer - c phase 4
Nodal depolarization occurs via what channel - Answer - c Ca2+ slow channel
Cardiac muscle cell depol occurs via what channel - Answer - c Na+ fast channel
During which phases is cell refractory - Answer - c 1,2,3
TDP caused by what mechanism usually - Answer - c EAD's (triggered activity)
Variance - Answer - c Take each number in sample, subtract each number from
mean, square each
difference, add all, divide sum by number in
sample -1
Standard deviation - Answer - c square root of variance
,P value - Answer - c Probability of certainty / smaller p value means the more likely
the result could
not occur by
chance
survival curve showing % of patients surviving treatment over time - Answer c
Kaplan - Meier
curves
What is 2 rules for CMC's like spiral or lasso - Answer - c 1. Only approved for
use in LA due to
chordae tendinea
2. Only rotate shaft clockwise
Stereotaxis catheters - Answer - c Have magnets in tip for maneuverability in
response to changes
in magnetic field
when measuring EGM's, earliest atrial activation usually from _____ and earliest
ventricular usually
from ______ - Answer - c P wave
/ QRS
Morady maneuver is used to differentiate - Answer - c differentiate AT from
AVNRT/AVRT
Most common arrhythmia - Answer - c AF
What system remotely steers guided catheters - Answer - c Hansen
robotic system ACT for using array - Answer - c ACT 250-300
bipolar cardiac electrodes record local _________ that occur during phase
_______ - Answer - c
, depolarizations /
phase 0
surface EKG leads usually filtered at what - Answer - c .1 - 100 Hz
Normal EGM filter settings - Answer - c 30 - 300 Hz
What to do if T waves interfere with A/V and why - Answer - c Increase the high
pass filter because
T waves are lower frequency as well as
far field
Unipolar filter setting - Answer - c .05 - 300 Hz
Best needle to use for pericardial space via subxyphoid approach - Answer - c
Tuohy needle
which type of guide wire is 260-300 cm long? - Answer - c exchange guide wires
(much longer than
diagnostic
catheter)
advantage of bipolar leads - Answer - c less prone to EMI noise
Unipolar leads more prone to - Answer - c 1. EMI / muscle artifact
oversensing
2. pectoralis muscle stimulation
(pocket stim)
Steroid tip leads reduce acute: - Answer - c inflammation and stimulation
threshold Best numbers for ventricular lead - Answer - c
How each is measured: Diagnostic catheter OD , Inflated balloon cath OD,
Needle OD, Guide wire -
Answer - c Diag: Fr / Balloon: mm / Needle: gauge / GW: inch
(thousandth of inches)
Standard sizes of micropuncture introducer sets and their purpose - Answer - c
21 gauge needle /
.018 inch wire. Purpose of micropuncture introducer set is for placement of
.035-.038 inch GW
Most common SVT - Answer - c AVNRT