The ASSE 6040 Standard applies to individuals who _______________ medical
gas and vacuum systems.
A install
B perform inspections of new installations
C maintain
D verify - CORRECT ANSWER ✔✔✔✔✔ maintain
At sea level the atmospheric pressure is 14.7 psia. What is the absolute pressure
of a 50 psig (gauge pressure) system at sea level?
A 64.7 psia
B 50 psia
C 34.7 psia
D 60 psia - CORRECT ANSWER ✔✔✔✔✔ 64.7 psia
Medical Gas _______________________________ shall be able to describe how to
find and implement manufacturer specific maintenance requirements for the
medical gas and vacuum system components per ASSE 6000 Annex D and CGA
E.10.
A Supplier or Distributor
B Maintenance Personnel
1
Page
EXAMPREPMASTER
,C Authority having jurisdiction
D Verifier or Inspector - CORRECT ANSWER ✔✔✔✔✔ Maintenance
Personnel
Medical Gas Maintenance Personnel shall be able to identify and describe the
hazards resulting from improper maintenance of medical gas and vacuum
systems, including:
A Flow
B Cross Connections
C Contaminants
D All of the above - CORRECT ANSWER ✔✔✔✔✔ All of the above
ASSE 6000 daily checks of bulk liquid systems include all of the following with
the exception of
______________.
A Primary and Reserve (if equipped) Tank Contents
B Pressure Dew Point Temperature
C Leaks or Unusual Icing
D Proper pressure maintained - CORRECT ANSWER ✔✔✔✔✔ Pressure Dew
Point Temperature
Per ASSE 6000 Annex D typical maintenance procedures include
_______________ checks of the Medical Air Moisture removal system (after
cooler, dryer drain, receiver drain, sight glass, etc.).
2
A Daily
Page
EXAMPREPMASTER
,B Weekly
C Monthly
D Annually - CORRECT ANSWER ✔✔✔✔✔ Daily
Per ASSE 6000 Annex D the Medical Air __________ is typically checked daily.
A Dew Point Monitor
B Carbon Monoxide Monitor
C Filter Condition
D Both A & B
E All of the above - CORRECT ANSWER ✔✔✔✔✔ Both A & B
Typical monthly checks, per ASSE 6000 Annex D, of area alarms include all of
the following except :
A Audible Signal
B Visual Signals
C Emergency Instructions
D Gauge Pressure Read Outs - CORRECT ANSWER ✔✔✔✔✔ Emergency
Instructions
Per ASSE 6000 Annex D, the medical air compressor intake location and filter
condition should be checked _______________.
A Daily
3
B Weekly
Page
EXAMPREPMASTER
, C Quarterly
D Annually - CORRECT ANSWER ✔✔✔✔✔ Quarterly
Minimum Maintenance Schedule, per ASSE 6000 Annex D, should include
annual checking of ________________ for Flow, Pressure/Vacuum, Damage, and
Wear.
A All outlets & inlets
B All critical care outlets & inlets
C 50% of all critical care outlets
D Annual testing of outlets & inlets is not recommended - CORRECT ANSWER
✔✔✔✔✔ All critical care outlets & inlets
Per ASSE 6000 Annex D the medical air compressor _____________________
should be calibrated annually.
A Dew Point & Carbon Monoxide Monitors
B Carbon Monoxide Monitor Only
C Gauges
D Motor Overloads - CORRECT ANSWER ✔✔✔✔✔ Dew Point & Carbon
Monoxide Monitors
Per ASSE 6000 Annex D & CGA E.10, after reassembly of equipment is shall be
tested prior to being returned to service for ______________________________.
A Gas Flow
B Pressure
4
Page
C Cross Connection
EXAMPREPMASTER