FRESENIUS VIRTUAL LEARNING PCT NEW HIRE
EXAM NEWEST 2025 ACTUAL EXAM COMPLETE
QUESTIONS AND CORRECT DETAILED ANSWERS
(VERIFIED ANSWERS) |ALREADY GRADED A+
Terms in this set (198)
The water system uses a sodium (Na+)
water softener to remove
calcium and magnesium.
When calcium and
magnesium are removed,
what is added to the water
in their place?
How often is water 2 times a day
hardness tested?
When is water hardness Before the first patient and at the end of the day
(caused by calcium and
magnesium) tested?
… 1/32
,3/27/25, 6:34 AM FRESENIUS VIRTUAL LEARNING PCT NEW HIRE EXAM NEWEST 2025 ACTUAL EXAM COMPLETE QUESTIONS AND CORR…
When the water is tested < 20 ppm (parts per million)
for water hardness
(caused by calcium and
magnesium) before the
first patient is put on AND
at the end of the day
(twice daily), what is the
acceptable amount of
these minerals found with
the SteriChek Water
Hardness Reagent Strips?
What test strips are used Water Check RC Strips
to measure residual
chlorine?
When do we use the after bleach day and before the first patient is put on
Water Check RC Strips
during the week?
If you find a vile of test no
strips that are not dated,
can we use them for
testing purposes?
… 2/32
,3/27/25, 6:34 AM FRESENIUS VIRTUAL LEARNING PCT NEW HIRE EXAM NEWEST 2025 ACTUAL EXAM COMPLETE QUESTIONS AND CORR…
What is the minimum 2
number for GAC (carbon
tanks) your clinic can ***Your clinic may have 4 GAC tanks
have?
In a clinic with 2 GAC worker, polisher
tanks, the first tank is
called the____ tank, and
the second tank is called
the ____ tank.
The worker tank is also primary
called the _____ tank.
The polisher tank is also secondary
called the _____ tank.
When the opener at your worker tank (primary tank)
clinic runs the RO for 15
minutes then waits the ***We are checking the functionality of the worker
required 2 minutes for the tank
sample valve to run
before collecting a
sample, after which GAC
tank is the valve located
for testing Total Chlorine?
How many employees 2 (opener plus a second employee)
have to watch the sample
for the Total Chlorine test ***One employee MUST be a RN
be obtained?
… 3/32
, 3/27/25, 6:34 AM FRESENIUS VIRTUAL LEARNING PCT NEW HIRE EXAM NEWEST 2025 ACTUAL EXAM COMPLETE QUESTIONS AND CORR…
Two employees (the 2
opener plus a second
employee) must be
present when the sample
for the Total Chlorine test
is obtained. How many
employees total must be
present when the test strip
is gently moved back and
forth in the sample cup?
How many times must the 3
sample cup for testing
Total Chlorine be rinsed ***Just like the Myron Meter
before obtaining the
sample?
How long does the test 60 seconds.
strip for Total Chlorine
need to be moved back
and forth in the water?
After the test strip for 20 seconds
Total Chlorine is moved
back and forth in the
water for 60 seconds, how
long do the two
employees have to wait
BEFORE reading the
results on the strip?
Does the Charge RN have yes
to be present during the
Total Chlorine test?
… 4/32