Answers 100% Pass
the process by which instruments and equipment are maintained to operate within
normal tolerance limits is called: - ✔✔Quality control
In order to insure the proper functioning of an automatic control system which of the
following must be evaluated at least once each year?
1. back up timer
2. minimum response time
3. density control functions - ✔✔1, 2, and 3
To insure that protective apparel is functioning properly protective aprons and gloves
should be... - ✔✔radiographically inspected at least once each year
an exposure of 600mA, 300 ms, and 89 kVp is used on a full wave rectified single-phase
x-ray unit. How many heat units will this exposure produce? - ✔✔16, 020 HU
100% Pass Guarantee Katelyn Whitman All Rights Reserved © 2025 1
, If a 180 cm SID to receptor distance is employed for a chest unit, to be in compliance the
light and radiation fields but coincide to within: - ✔✔3.6 cm on each border of the
image receptor
which of the following devices is used in a quality control program to check the
exposure rate of a radiographic or fluoroscopic unit? - ✔✔mR / mAs meter
the radiology department plans on changing the film - screen imaging system it
currently employs. Which of the following testing procedures would be an essential
part in this selection process? - ✔✔sensitometry
the SID indicator on all new radiographic units must be accurate to within... - ✔✔2% of
the actual measurable SID
the instantaneous loading capacity rating of the anode is primarily effected by the: -
✔✔area of the focal tract
a full wave single-phase rectified radiographic exposure produces 900 HU in the anode.
Using the same exposure factor on a 3-phase 6-pulse unit would be associated with the
production of: - ✔✔1215 HU
which type of equipment evaluation is required for all new pieces of radiographic units
prior to being placed into service? - ✔✔acceptance testing
100% Pass Guarantee Katelyn Whitman All Rights Reserved © 2025 2