1. The primary responsibility of a medical gas verifier is to
n n n n n n n n n n
A. Install medical gas piping n n n
B. Verify that medical gas and vacuum systems are installed and f
n n n n n n n n n n
unctioning correctly n
C. Maintain patient monitoring equipment n n n
D. Operate medical gas alarms n n n
Verification ensures safety, accuracy, and compliance before system
n n n n n n n
s are placed into service.
n n n n
2. Medical gas verification is required before a system is
n n n n n n n n
A. Pressure tested n
B. Labeled
C. Used for patient care n n n
D. Designed
Verification confirms systems are safe prior to clinical use.
n n n n n n n n
3. Which standard is most commonly referenced for medical gas v
n n n n n n n n n
erification in New Jersey? n n n
A. OSHA 1910 n
B. NFPA 99 n
C. NFPA 13 n
, D. ASME Section IX n n
NFPA 99 governs Health Care Facilities Code requirements.
n n n n n n n
4. Who is permitted to perform medical gas verification?
n n n n n n n
A. Any licensed plumber
n n
B. Hospital maintenance staff n n
C. A qualified and credentialed medical gas verifier
n n n n n n
D. The project architect
n n
Verification must be done by an independent qualified individual.
n n n n n n n n
5. Medical gas verification must be independent from
n n n n n n
A. The hospital administration
n n
B. The inspection agency
n n
C. The installer of the system
n n n n
D. The local authority
n n
Independence avoids conflicts of interest. n n n n
6. Cross-connection testing ensures that n n n
A. Alarms function properly n n
B. Pressure is correct n n
C. Each outlet delivers only the intended gas
n n n n n n
D. Valves are accessible n n
This prevents delivery of the wrong gas to patients.
n n n n n n n n
7. Which gas is commonly color-coded green in medical gas systems?
n n n n n n n n n
A. Nitrous oxide n
B. Medical air n
C. Vacuum
D. Oxygen
Color coding helps prevent gas identification errors.
n n n n n n
, 8. A pressure drop test is used to verify
n n n n n n n
A. Flow rate accuracy
n n
B. Alarm sensitivity n
C. System tightness and absence of leaks
n n n n n
D. Outlet labeling n
Pressure stability indicates leak-free piping.
n n n n
9. Medical vacuum systems are tested primarily for
n n n n n n
A. Pressure
B. Purity
C. Flow and suction levels
n n n
D. Odor
Vacuum performance is measured by suction capability.
n n n n n n
10. Which component must be verified for proper alarm
n n n n n n n n
activation?
A. Zone valves
n
B. Master alarm panels n n
C. Outlet plates n
D. Pipe supports
n
Alarms are critical for alerting staff to system issues.
n n n n n n n n
11. Gas identity testing confirms
n n n
A. Pipe material
n
B. Pressure class n
C. Correct gas at each outletn n n n
D. Valve locationn
Gas identity errors can be fatal.
n n n n n
12. A verifier must document
n n n
A. Only failed tests
n n