TEST EXAM QUESTIONS AND
ANSWERS
, B. Authority having jurisdiction - ANSWERSAn existing system that is not in strict
compliance with the provisions of this code shall be permitted to be continued in a use
as long as the __________ has determined that such use does not constitute a distinct
hazard to life.
A. Architect
B. Authority having jurisdiction
C. Engineer
D. Nursing
B. Designated storage area - ANSWERSIndoor locations for oxygen, nitrous oxide, and
mixtures of these gases shall not communicate with any of the following
except__________
A. Areas with open flames
B. Designated storage area
C. Engines
D. Kitchens
D. Visual pressure indicator - ANSWERSAn auxiliary source connection shall consist of
all the following except
A. Removable plug or cap
B. Tree
C. Valve
D. Visual pressure indicator
A. Each building served - ANSWERSWhen multiple buildings are served from a single
cryogenic liquid oxygen source, it is required to place the emergency oxygen supply
connection in-
A. Each building served
B. The building closest to the source
C. The largest building served
D. The main building only
B. 3 - ANSWERSThe minimum clearance around the EOSC for a temporary source of
supply is ____________ ft.
A. 2
B. 3
C. 5
D. 6
A. Except be filtered to 0.01 microns - ANSWERSThe quality medical air shall comply
with all of the following standards except
A. Be filtered to 0.01 microns