AASM scoring rules and guidelines
questions well solved
Maximum electrode impedances - correct answer ✔✔5 KΩ
Minimum registry length or digital amplitude resolution - correct answer ✔✔12 bits per sample
EEG sampling rates - correct answer ✔✔Max: 500Hz
Min: 200 Hz
EOG sampling rates - correct answer ✔✔Max: 500 Hz
Min: 200Hz
EMG sampling rates - correct answer ✔✔Max: 500 Hz
Min: 200 Hz
ECG sampling rates - correct answer ✔✔Max: 500 Hz
Min: 200 Hz
Airflow sampling rate - correct answer ✔✔Max: 100 Hz
Min: 25Hz
Oximetry/trans cutaneous PCO2 sampling rate - correct answer ✔✔Max: 25 Hz
Min: 10 Hz
,Nasal Pressure, End-Tidal PCO2, PAP device flow sampling rates - correct answer ✔✔Max: 100
Hz
Min: 25 Hz
Esophageal pressure sampling rates - correct answer ✔✔Max: 100 Hz
Min: 25 Hz
Body position sampling rates - correct answer ✔✔Max: 1 Hz
Min: 1 Hz
Snoring sounds sampling rates - correct answer ✔✔Max: 500 Hz
Min: 200 Hz
Rib cage and Abdominal movements sampling rates - correct answer ✔✔Max: 100 Hz
Min: 25 Hz
EEG Filter settings - correct answer ✔✔LFF: 0.3 Hz
HFF: 35 Hz
EOG filter settings - correct answer ✔✔LFF: 0.3 Hz
HFF: 35 Hz
EMG filter settings - correct answer ✔✔LFF: 10 Hz
HFF: 100 Hz
ECG filter settings - correct answer ✔✔LFF: 0.3 Hz
, HFF: 100 Hz
Oronasal thermal flow, Thoracoabdominal belts filter settings - correct answer ✔✔LFF: 0.1 Hz
HFF: 15 Hz
Nasal Pressure filter settings - correct answer ✔✔LFF: Direct current (DC channel) or </= 0.03
Hz
HFF: 100 Hz
PAP device flow filter settings - correct answer ✔✔LFF: DC channel
HFF: DC channel
Snoring filter settings - correct answer ✔✔LFF: 10 Hz
HFF: 100 Hz
The recommended EEG derivations are: - correct answer ✔✔F4-M1
C4-M1
O2-M1
What is the standard reference electrode for recording EEG? - correct answer ✔✔M1
If M1 fails, what is the backup electrode you would use? - correct answer ✔✔M2
The recommended EOG derivations are: - correct answer ✔✔E1-M2 and E2-M2
questions well solved
Maximum electrode impedances - correct answer ✔✔5 KΩ
Minimum registry length or digital amplitude resolution - correct answer ✔✔12 bits per sample
EEG sampling rates - correct answer ✔✔Max: 500Hz
Min: 200 Hz
EOG sampling rates - correct answer ✔✔Max: 500 Hz
Min: 200Hz
EMG sampling rates - correct answer ✔✔Max: 500 Hz
Min: 200 Hz
ECG sampling rates - correct answer ✔✔Max: 500 Hz
Min: 200 Hz
Airflow sampling rate - correct answer ✔✔Max: 100 Hz
Min: 25Hz
Oximetry/trans cutaneous PCO2 sampling rate - correct answer ✔✔Max: 25 Hz
Min: 10 Hz
,Nasal Pressure, End-Tidal PCO2, PAP device flow sampling rates - correct answer ✔✔Max: 100
Hz
Min: 25 Hz
Esophageal pressure sampling rates - correct answer ✔✔Max: 100 Hz
Min: 25 Hz
Body position sampling rates - correct answer ✔✔Max: 1 Hz
Min: 1 Hz
Snoring sounds sampling rates - correct answer ✔✔Max: 500 Hz
Min: 200 Hz
Rib cage and Abdominal movements sampling rates - correct answer ✔✔Max: 100 Hz
Min: 25 Hz
EEG Filter settings - correct answer ✔✔LFF: 0.3 Hz
HFF: 35 Hz
EOG filter settings - correct answer ✔✔LFF: 0.3 Hz
HFF: 35 Hz
EMG filter settings - correct answer ✔✔LFF: 10 Hz
HFF: 100 Hz
ECG filter settings - correct answer ✔✔LFF: 0.3 Hz
, HFF: 100 Hz
Oronasal thermal flow, Thoracoabdominal belts filter settings - correct answer ✔✔LFF: 0.1 Hz
HFF: 15 Hz
Nasal Pressure filter settings - correct answer ✔✔LFF: Direct current (DC channel) or </= 0.03
Hz
HFF: 100 Hz
PAP device flow filter settings - correct answer ✔✔LFF: DC channel
HFF: DC channel
Snoring filter settings - correct answer ✔✔LFF: 10 Hz
HFF: 100 Hz
The recommended EEG derivations are: - correct answer ✔✔F4-M1
C4-M1
O2-M1
What is the standard reference electrode for recording EEG? - correct answer ✔✔M1
If M1 fails, what is the backup electrode you would use? - correct answer ✔✔M2
The recommended EOG derivations are: - correct answer ✔✔E1-M2 and E2-M2