Correct A+ Pass
What 2 things can be used to protect against neutron leakage from the treatment room
door? - Answer- 1. Maze-increase the length (>5 meters)
2. Material-add few inches of hydrogenous material such as polyethylene
3 organizations that are available for accreditation for radiation oncology practices -
Answer- 1. American College of Radiology (ACR)
2. American College of Radiation Oncology (ACRO)
3. American Society for Radiation Oncology (ASTRO)
What can be added to a treatment room door to protect against scattered x-rays? -
Answer- Steel or lead sheet
What can be added to a treatment room door to thermalize neutrons and reduce the
neutron dose? - Answer- Few inches of a hydrogenous material such as polyethylene
What is the name of the ACR's accreditation program? - Answer- ACR's Radiation
Oncology Practice Accreditation Program (ROPA)
What is the name of the accrediting program for radiation oncology by ASTRO? -
Answer- Apex-Astro Accreditation Program for Excellence
What are the 3 types of radiation that protection is required for? - Answer- 1. Primary
radiation
2. Scattered radiation
3. Leakage radiation through the source housing
List the 4 factors that are used in the calculation of barrier thickness - Answer- 1.
Workload
2. Use factor
3. Occupancy factor
4. Distance
What are the 4 factors that the amount of radiation scattered depends on? - Answer- 1.
Beam intensity incident on the scatterer
2. The quality of radiation
3. The area of the beam of the scatterer
4. The scattering angle
What document outlines the leakage requirement for therapeutic source assemblies in
radiation therapy? What was the old document that it replaced? - Answer- 1. NCRP
Report 102
, 2. NCRP Report 33
For door shielding-what are the requirements? - Answer- Unless a maze entranceway is
provided, the door must provide shielding equivalent to the wall surrounding the door,
be extremely heavy, requires a motor drive as a means of manual operation in case of
emergency.
What are the high energy x-ray beams (>10 MV) contaminated with? What energy
range dose this increase rapidly? - Answer- Neutrons, 10-20MV
What is sufficient for protection of neutrons? - Answer- Concrete barriers designed for x-
ray shielding
What method was traditionally used for treatment verification? - Answer- Port films
What has typically replaced the traditional method for treatment verification? - Answer-
Using electronic portal imaging devices (EPIDS)
ACR is the primary professional organization for who? - Answer- Radiologists, radiation
oncologists, and clinical medical physicists
Items that fall under the realm of the NRC are: - Answer- 1. Use of radioisotopes in
brachytherapy
2. Cobalt source
What is the latest equipment available to check patient set-up immediately prior to
treatment and to adjust patient position as needed to localize the target volume
precisely? - Answer- IGRT
What is necessary to have when checking patient set-up prior to treatment to localize
the target volume precisely? - Answer- Reference images for comparison
Who is responsible for the verification and documentation of the accuracy of the
treatment delivery as related to the initial treatment planning & set up procedure? -
Answer- Physician
What does OSHA do? - Answer- 1. Mandated a policy on bloodborne pathogens
2. Sets standards for exposure to cadmium and lead
Additional practices that can facilitate a positive culture shift toward safety mindfulness
examples: (3) - Answer- 1. Good catch
2. Safety rounds
3. Huddles
What is the recommended frequency of testing for the ODI? - Answer- Daily