RPSGT PRACTICE EXAM
Study online at https://quizlet.com/_h17otj
1. A PATIENTS OXIMETER READING PLACE THE PROBE ON THE PATIENTS EARLOBE
DURING PATIENT CALIBRATION
IS 88%. WHICH OF THE FOLLOW-
ING WOULD BE THE APPROPRI-
ATE RESPONSE?
2. A PATIENT WITH ONDINES CENTRAL SLEEP APNEA
CURSE WOULD BE EXPECTED TO
HAVE WHICH TYPE OF RESPIRA-
TORY IMPAIRMENT IN SLEEP
3. ACCORDING TO THE AASM, IN- DETECTION OF RESPIRATORY EFFORT
TERCOSTAL EMG ACTIVITY IS
AN ACCEPTABLE ALTERNATIVE
SESOR FOR RECORDING WHAT
PARMETER
4. SLOW ROLLING EYE MOVE- 0.3HZ
MENTS OCCURING DURING
STAGE N1 SLEEP WILL BEST BE
SEEN AT LOW FILTER SETTING OF
5. A 1.0 HZ EEG WAVE IS MEASURED 75UV
AT 15MM PEAK TO PEAK. THE
SENSITIVITY IS 5UV/MM. THE
VOLTAGE OF THE WAVE IS
6. A CHEYNE STOKES BREATHING CONGESTIVE HEART FAILURE
PATTERN IS OFTEN SEEN IN PA-
TIENTS WITH
7. CAPNOGRAPHY MEASURES EXPIRED CO2
, RPSGT PRACTICE EXAM
Study online at https://quizlet.com/_h17otj
8. WHICH OF THE FOLLOWING 100HZ
HIGH FREQUENCY FILTER SET-
TINGS IS RECOMMENDED FOR
RECORDING SUBMENTAL EMG
9. ENDOESOPHAGEAL PRESSURE SLEEP RELATED BREATHING DISORDERS
MEASUREMENTS WOULD BE
MOST HELPFUL IN EVALUTATING
WHICH SLEEP DISORDERS
10. A PATIENT BEING EVALUATED CONTINUE TO OBSERVE AND DOCUMENT RHYTHM
FOR OBSTRUCTIVE SLEEP DIS-
ORDERED BREATHING HAS ATRI-
AL FIBRILLATION. WHAT SHOULD
THE TECH DO
11. WHICH OF THE FOLLOWING IS VENTRICULAR FIBRILLATION
THE MOST DANGEROUS CAR-
DIAC RHYTHM
12. HOW WOULD YOU COMPUTE TOTAL SLEEP TIME DIVIDED BY TIME IN BED X 100
SLEEP EFFICIENCY
13. HOW DO YOU CALCULATE APNEA # APNEAS AND HYPOPNEAS DIVIDED BY TST X60
+ HYPOPNEA INDEX
14. WHICH HIGH PASS FILTER SET- 0.1HZ
TING IS MOST APPROPRIATE FOR
RECORDING RESPIRATION
15. THE PURPOSE OF AN EXHALA- TO VENT EXHALED C02
TION PORT IN A CPAP CIRCUIT IS
16. THE QRS IS WIDE AND BIZARRE
Study online at https://quizlet.com/_h17otj
1. A PATIENTS OXIMETER READING PLACE THE PROBE ON THE PATIENTS EARLOBE
DURING PATIENT CALIBRATION
IS 88%. WHICH OF THE FOLLOW-
ING WOULD BE THE APPROPRI-
ATE RESPONSE?
2. A PATIENT WITH ONDINES CENTRAL SLEEP APNEA
CURSE WOULD BE EXPECTED TO
HAVE WHICH TYPE OF RESPIRA-
TORY IMPAIRMENT IN SLEEP
3. ACCORDING TO THE AASM, IN- DETECTION OF RESPIRATORY EFFORT
TERCOSTAL EMG ACTIVITY IS
AN ACCEPTABLE ALTERNATIVE
SESOR FOR RECORDING WHAT
PARMETER
4. SLOW ROLLING EYE MOVE- 0.3HZ
MENTS OCCURING DURING
STAGE N1 SLEEP WILL BEST BE
SEEN AT LOW FILTER SETTING OF
5. A 1.0 HZ EEG WAVE IS MEASURED 75UV
AT 15MM PEAK TO PEAK. THE
SENSITIVITY IS 5UV/MM. THE
VOLTAGE OF THE WAVE IS
6. A CHEYNE STOKES BREATHING CONGESTIVE HEART FAILURE
PATTERN IS OFTEN SEEN IN PA-
TIENTS WITH
7. CAPNOGRAPHY MEASURES EXPIRED CO2
, RPSGT PRACTICE EXAM
Study online at https://quizlet.com/_h17otj
8. WHICH OF THE FOLLOWING 100HZ
HIGH FREQUENCY FILTER SET-
TINGS IS RECOMMENDED FOR
RECORDING SUBMENTAL EMG
9. ENDOESOPHAGEAL PRESSURE SLEEP RELATED BREATHING DISORDERS
MEASUREMENTS WOULD BE
MOST HELPFUL IN EVALUTATING
WHICH SLEEP DISORDERS
10. A PATIENT BEING EVALUATED CONTINUE TO OBSERVE AND DOCUMENT RHYTHM
FOR OBSTRUCTIVE SLEEP DIS-
ORDERED BREATHING HAS ATRI-
AL FIBRILLATION. WHAT SHOULD
THE TECH DO
11. WHICH OF THE FOLLOWING IS VENTRICULAR FIBRILLATION
THE MOST DANGEROUS CAR-
DIAC RHYTHM
12. HOW WOULD YOU COMPUTE TOTAL SLEEP TIME DIVIDED BY TIME IN BED X 100
SLEEP EFFICIENCY
13. HOW DO YOU CALCULATE APNEA # APNEAS AND HYPOPNEAS DIVIDED BY TST X60
+ HYPOPNEA INDEX
14. WHICH HIGH PASS FILTER SET- 0.1HZ
TING IS MOST APPROPRIATE FOR
RECORDING RESPIRATION
15. THE PURPOSE OF AN EXHALA- TO VENT EXHALED C02
TION PORT IN A CPAP CIRCUIT IS
16. THE QRS IS WIDE AND BIZARRE