What are the three main maximum exposures? - Answer Main magnetic field
Gradient magnetic field
Radio frequency power (RF)
Possible effects of the main magnetic field? - Answer tissue heating and cardiac
composition
Possible effects of the gradient magnetic field? - Answer no more than 6T/sec
Possible effects: magneto-phosphenes, bones heating faster, cardiac fibrillation
Requirement level for safety with the radio frequency? - Answer the whole body - no
more than 4.0 w/kg
the head - no more than 3.2 w/kg
Where is most of the RF power used in MR imaging? - Answer transformed into heat that
is absorbed in the patient's tissues
Bio effect of the RF absorption - Answer heating of the tissue
Limit to the increases in core body temperature by the FDA - Answer 1C
Are there known long-term adverse biological effects associated with extended
exposure to the magnetic fields used in MRI - Answer No
Guidelines for evaluating electromagnetic exposure risks for trials of clinical MRI are
issued by? - Answer FDA
The Blue-Ribbon Panel's Mission - Answer to produce guidelines for MRI safety entitled
ACR Guidance Document for Safe MRI Practices
White Paper on MRI Safety (2020) - Answer intended to provide guidelines for MRI
facilities for the development and implementation of safety policies and procedures
Hardwear for image creation that must be used to test implants or devices? - Answer
static main magnetic field, RF field, gradient field
Terminology for MRI safety - Answer "MRI safe"
"MRI unsafe"
"MRI conditional"
, The primary effects of the static magnetic field during a MR scan? - Answer implants,
devices, projectiles
Gradients and RF fields are considered TVMF, what is TVMF - Answer Time Varying
Magnetic Fields
MRI conditional has how many subcategories? - Answer 8
MRI unsafe is subdivided into categories - Answer MRI unsafe 1, MRI unsafe 2
The bio-effects of RF irradiation - Answer Tissue heating, RF antennae effects (implant
heats up), Thermal injuries
The greatest concern for RF effects are - Answer Fast spin echos
What happens in spin echo? - Answer at least one addition RF pulse is applied
What is the method to control RF absorption? - Answer SAR - specific absorption rate
What does SAR stand for? What does it do? - Answer SAR - specific abortion rate.
Is the method of control RF absorption.
The biologic effect of RF absorption - Answer tissue heating
Why is SAR monitored? - Answer To help monitor heating of patient tissue
RF characteristics to calculate SAR levels - Answer induced electric field and pulse duty
cycle
Higher energy pulse more likely it will induce electrical current
Patient characteristics to calculate SAR levels - Answer tissue density, conductivity,
patient's size
FDA limit for temperature is an increase of ___ in the core of the body ___ in the head ___
in the trunk ___ in the extremities - Answer 1.0C
38C body
39C head
40C extremities
Coupling transmitters to receivers can be a cause of what type of injury? - Answer
Thermal
Equipment like coils' cords, ECG leads must be _____ to prevent in injuires. - Answer
insulated and properly used
Used to spatially encode signal in three planes and generate echoes - Answer
GRADIENT MAGNETIC FIELDS
Solution
What are the three main maximum exposures? - Answer Main magnetic field
Gradient magnetic field
Radio frequency power (RF)
Possible effects of the main magnetic field? - Answer tissue heating and cardiac
composition
Possible effects of the gradient magnetic field? - Answer no more than 6T/sec
Possible effects: magneto-phosphenes, bones heating faster, cardiac fibrillation
Requirement level for safety with the radio frequency? - Answer the whole body - no
more than 4.0 w/kg
the head - no more than 3.2 w/kg
Where is most of the RF power used in MR imaging? - Answer transformed into heat that
is absorbed in the patient's tissues
Bio effect of the RF absorption - Answer heating of the tissue
Limit to the increases in core body temperature by the FDA - Answer 1C
Are there known long-term adverse biological effects associated with extended
exposure to the magnetic fields used in MRI - Answer No
Guidelines for evaluating electromagnetic exposure risks for trials of clinical MRI are
issued by? - Answer FDA
The Blue-Ribbon Panel's Mission - Answer to produce guidelines for MRI safety entitled
ACR Guidance Document for Safe MRI Practices
White Paper on MRI Safety (2020) - Answer intended to provide guidelines for MRI
facilities for the development and implementation of safety policies and procedures
Hardwear for image creation that must be used to test implants or devices? - Answer
static main magnetic field, RF field, gradient field
Terminology for MRI safety - Answer "MRI safe"
"MRI unsafe"
"MRI conditional"
, The primary effects of the static magnetic field during a MR scan? - Answer implants,
devices, projectiles
Gradients and RF fields are considered TVMF, what is TVMF - Answer Time Varying
Magnetic Fields
MRI conditional has how many subcategories? - Answer 8
MRI unsafe is subdivided into categories - Answer MRI unsafe 1, MRI unsafe 2
The bio-effects of RF irradiation - Answer Tissue heating, RF antennae effects (implant
heats up), Thermal injuries
The greatest concern for RF effects are - Answer Fast spin echos
What happens in spin echo? - Answer at least one addition RF pulse is applied
What is the method to control RF absorption? - Answer SAR - specific absorption rate
What does SAR stand for? What does it do? - Answer SAR - specific abortion rate.
Is the method of control RF absorption.
The biologic effect of RF absorption - Answer tissue heating
Why is SAR monitored? - Answer To help monitor heating of patient tissue
RF characteristics to calculate SAR levels - Answer induced electric field and pulse duty
cycle
Higher energy pulse more likely it will induce electrical current
Patient characteristics to calculate SAR levels - Answer tissue density, conductivity,
patient's size
FDA limit for temperature is an increase of ___ in the core of the body ___ in the head ___
in the trunk ___ in the extremities - Answer 1.0C
38C body
39C head
40C extremities
Coupling transmitters to receivers can be a cause of what type of injury? - Answer
Thermal
Equipment like coils' cords, ECG leads must be _____ to prevent in injuires. - Answer
insulated and properly used
Used to spatially encode signal in three planes and generate echoes - Answer
GRADIENT MAGNETIC FIELDS