CCHT water testing questions with correct answers
Chlorine testing correct answer reverse osmosis (ro) machines runs 15 minutes before performing any total chlorine testing 100 ml correct answer the correct amount of water to collect for total chlorine testing Rpc ultra-low total chlorine strips correct answer strips used for chlorine testing How long are rpc strips correct answer immersed in sample water for 60 seconds When should chlorine be tested correct answer prior to treatments in the morning and every 3-4 hours while patients are dialyzing Rpc ultra -low total chlorine strips test correct answer chlorine in the water after carbon tank and post ro 0.00 to 0.09 ppm correct answer total chlorine levels that are patient safe and document in "tms" or paper log First sample correct answer should be taken from post carbon tank 1 (worker) Second sample correct answer post ro If chlorine is noted post ro correct answer do not initiate any patient treatment, and document per facility policy and procedure Chlorine breakthrough correct answer occurs post carbon tank 1 (worker), immediate additional testing should take place after carbon tank 2 (polisher)