Sterile Processing Certification Questions And Answers Complete Solution
Sterile Processing Certification Questions And Answers Complete Solution What is the area temp and humidity for deco tam? 60-65 degF 30-60% humidity What is the air exchange rate and pressure type for decontam? 10 air exchanges per hour under negative pressure What is the area temp and humidity for the sterile storage area? 68-75 degF 35-70% humidity What is the area temp and humidity for the prep and packaging area? 68-73 degF ideal humidity 50% no lower than the 35% The main role of sterile processing is to: provide dependable reliable services to enhance the quality of patient care Ethics is the practice of: doing the right thing What established sterilization practice standards based on opinion, research, and/or theory? AAMI (Association for the Advancement of Medical Instrumentation)