MAINTENANCE, AND SAFETY UPDATED ACTUAL QUESTIONS
AND CORRECT ANSWERS
Question:
1. Nova StatStrip
Answer:
A glucometer used to measure blood glucose (blood sugar).
Question:
2. Certification
Answer:
You must be certified to operate the meter, and certification expires yearly.
Question:
3. Certification maintenance
Answer:
Scan your ID, run Level 1 (LO) and Level 2 (HI) controls, and demonstrate competency annually.
Question:
4. Operator badge
Answer:
Never borrow or loan your badge.
Question:
5. Patient information
Answer:
Never keep patient information in the meter carrying case.
Question:
6. Meter temperature
Answer:
15-40°C.
Question:
7. Opened test strips
Answer:
Good for 180 days (6 months).
Question:
8. Opened control solution
Answer:
Good for 90 days (3 months).
, Question:
9. Test strip storage Keep at room temperature and never freeze.
Answer:
Question:
10. Cleaning the meter
Answer:
Clean after every patient.
Question:
11. Cleaning supplies 1:10 bleach, 70% isopropanol, or an approved disinfectant wipe.
Answer:
Question:
12. Cleaning safety
Answer:
Never immerse, run water over, or spray disinfectant directly on the meter.
Question:
13. Capillary blood
Answer:
Test immediately.
Question:
14. Venous or arterial blood
Answer:
Test within 30 minutes.
Question:
15. Recommended anticoagulant
Answer:
Heparin.
Question:
16. Required blood sample
Answer:
1.2 μL.
Question:
17. Poor circulation
Answer:
Can make capillary blood results unreliable.
Question:
18. Hypoglycemia