MDCB CERTIFICATION MATH |ACTUAL QUESTIONS
AND VERIFIED ANSWERS|BRAND NEW 2026-2027
UPDATE|GRADED A+
Question 1
If the exposure rate constant is 0.87 Rcm2hr-1mCi-1 and the average patient transmission
factor
is 0.20, the exposure rate at 12.5 cm for a patient who has been injected with 20 mCi of
Tc99m
is:
(A) 18 mR/hr.
(B) 20 mR/hr.
(C) 22 mR/hr.
(D) 25 mR/hr.
CORRECT ANSWER
(C) 22 mR/hr.
0.87x20x0.2x(1/12.5)^2=0.022R or 22mR
Question 2
The exposure rate emitting from a brachytherapy safe is found to be 25 mR/hr at 1 m from
the
safe source. The minimum thickness of shielding necessary to reduce the exposure to an
adjacent office secretary to an acceptable level would be:
(A) 1 TVL.
(B) 2 TVL.
(C) 10 TVL.
1
, (D) 12 TVL.
CORRECT ANSWER
(B) 2 TVL.
Question 3
The mass attenuation coefficient for lead at 140 keV is 2.6 cm2/g and the density of lead is
11.3 g/cm3. What is the percentage transmission through a 0.5 mm thick lead apron for an
effective energy of 140 keV?
(A) 11.3%
(B) 23.0%
(C) 32.3%
(D) 43.7%
CORRECT ANSWER
(B) 23.0%
2.6/11.3=0.23
Question 4
The dose rate at a patient's mid-plane is found to be 250 cGy/min at 100 cm SAD. A
protocol for hemi-body irradiation stipulates that the dose rate must be no more than 100
cGy/min. At what minimum SAD must the patient be treated?
(A) 120 cm
(B) 145 cm
(C) 158 cm
(D) 182 cm
CORRECT ANSWER
(C) 158 cm
2
AND VERIFIED ANSWERS|BRAND NEW 2026-2027
UPDATE|GRADED A+
Question 1
If the exposure rate constant is 0.87 Rcm2hr-1mCi-1 and the average patient transmission
factor
is 0.20, the exposure rate at 12.5 cm for a patient who has been injected with 20 mCi of
Tc99m
is:
(A) 18 mR/hr.
(B) 20 mR/hr.
(C) 22 mR/hr.
(D) 25 mR/hr.
CORRECT ANSWER
(C) 22 mR/hr.
0.87x20x0.2x(1/12.5)^2=0.022R or 22mR
Question 2
The exposure rate emitting from a brachytherapy safe is found to be 25 mR/hr at 1 m from
the
safe source. The minimum thickness of shielding necessary to reduce the exposure to an
adjacent office secretary to an acceptable level would be:
(A) 1 TVL.
(B) 2 TVL.
(C) 10 TVL.
1
, (D) 12 TVL.
CORRECT ANSWER
(B) 2 TVL.
Question 3
The mass attenuation coefficient for lead at 140 keV is 2.6 cm2/g and the density of lead is
11.3 g/cm3. What is the percentage transmission through a 0.5 mm thick lead apron for an
effective energy of 140 keV?
(A) 11.3%
(B) 23.0%
(C) 32.3%
(D) 43.7%
CORRECT ANSWER
(B) 23.0%
2.6/11.3=0.23
Question 4
The dose rate at a patient's mid-plane is found to be 250 cGy/min at 100 cm SAD. A
protocol for hemi-body irradiation stipulates that the dose rate must be no more than 100
cGy/min. At what minimum SAD must the patient be treated?
(A) 120 cm
(B) 145 cm
(C) 158 cm
(D) 182 cm
CORRECT ANSWER
(C) 158 cm
2