A 64-slice MSCT system employs an array of 64 detectors, each with a dimension
of 0.625 mm. What beam collimation is required to expose the middle 32
detectors of the array to transmitted x-radiation?
A. 10 mm
B. 20 mm
C. 32 mm
D. 64 mm correct answer B
An acquisition is made on a 4-slice MSCT system with a detector array of 16 x
1.25-mm detector elements along the z-axis. With a selected beam width of 20
mm, what beam pitch would result in the table moving 35.00 mm for each
rotation of the gantry?
A. 1.0
B. 1.5
C. 1.75
D. 2.0 correct answer C