Answers Latest Updated
The central ray indicator must align to the actual center of the x-ray field with a
tolerance of: - answer1% source-to-image distance (SID)
The Source-to-Image Distance (SID) indicator misrepresent, the actual SID within a
tolerance of: - answer+- 2% of the SID
The light field must correspond to the actual exposure field within a tolerance of: -
answer+- 2% of the SID
" Coincidence testing" of the x-ray to assembly is an evaluation of which of the
following? - answer Light field accuracy
When the source to image distance (SID) is 80cm (31 inches), the actual century should
be within what distance of the indicated x-ray field center? - answer0.8 cm (0.3 inches)
At a source to image distance (SID) of 60 inches (152 cm), the central ray must align to
the central ray indicator within: - answer0.6 inch (1.5cm)
What is the primary purpose of regularly testing in calibrating the central ray alignment?
- answerMaintain the accuracy of the radiographic projection
When the source the image distance (SID) is 80cm (31.5 inches), the actual SID should
match the displayed SID within a range of: - answer+-1.6 cm (0.6 inches)
When the source the image distance (SID) is 80 cm (31.5 inches), the Light field should
match the exposure field within a range of: - answer+-1.6 cm (0.6 inches)
When the source the image distance (SID) is 40 inches (102 cm), the actual SID should
match the displayed SID within a range of: - answer+- 0.8 inches (4.1 cm)
Actual exposure time must be within what percentage of indicated exposure time? -
answer+-5%
The actual exposure time must be accurate within what range when the assigned
exposure time is 80 milliseconds (ms)? - answer+-4.0 ms
The actual exposure time must be accurate within what range when the assigned
exposure time is 100 milliseconds (ms) - answer+-5 ms
, The actual exposure time must be accurate within what range when the assigned
exposure time is 72 milliseconds (ms) - answer+-3.6 ms
If the assigned exposure time is 60 milliseconds (ms), the actual exposure time should
be no more than: - answer63.0 ms
Which of these exposure times is outside acceptable limits when the exposure time on
the operators consul, is set to 300 milliseconds (ms) - answer316 ms
Consider this scenario: a medical physicist is testing an automatic exposure control
(AEC) system. If an exposure through 1 cm of aluminum results in a receptor exposure
of 10 micro gray, and exposure through 2 centimeters of aluminum should result in a
receptor exposure of: - answer10 microgray (ugy)
Consider this scenario: a medical physicist is testing an automatic exposure control
(AEC) system. If an exposure through 10 cm of aluminum results in a receptor exposure
of 100 micro gray, and exposure through 5 centimeters of aluminum should result in a
receptor exposure of: - answer100 microgray (ugy)
A radiograph is acquired using an automatic exposure controlled (AEC) unit with an
exposure setting of 200 milliamperage (mA) and 60 Kilovoltage peak (kVp). If the initial
receptor exposure is measured as 20 micro gray (Ugy), increasing the technique to 69
kVp will result in a receptor exposure of: - answer20 uGy
A radiograph is acquired using an automatic exposure controlled (AEC) unit with an
exposure setting of 200 milliamperage (mA) and 100 Kilovoltage peak (kVp). If the initial
receptor exposure is measured as 50 micro gray (Ugy), increasing the technique to 300
kVp will result in a receptor exposure of: - answer50 Ugy
the actual Kilovoltage peak (kVp) of an X-ray exposure must correspond to the selected
kVp within a tolerance of what percentage? - answer+-5%
X-ray systems must demonstrate consistent exposure reproducibility within a tolerance
of what percentage - answer+-5%
The X-ray tube output for identical technical parameters should not vary by more than
what percentage? - answer+-5%
The milligray/milliampere-seconds (mGy/mAs) measurement between adjacent mA
stations should not vary by more than: - answer+-10%
X-ray systems must demonstrate exposure linearity between adjacent milliamperage
(mA) stations within a tolerance of what percentage? - answer+-10%