CRCST Exam Prep Practice
Exercise 1 Questions and Answers
100% Correct
Dynamic air removal: exposure time for any wrapped item at 275 degrees. -
ANSWER-3 minutes
What is the correct percentage of relative humidity for steam sterilization? -
ANSWER-97%-100%
What are the 4 cycles of steam sterilization? - ANSWER-Conditioning,
exposure, exhaust, drying
Which low-temp sterilization method requires monitoring? - ANSWER-
Ethylene oxide
, How does ethylene oxide kill microbes? - ANSWER-By a process called
alkylation.
How do ozone and hydrogen peroxide kill microbes? - ANSWER-By a process
called oxidation.
What are the 4 cycles for vaporized hydrogen peroxide systems? - ANSWER-
Conditioning, leak test, sterilization, aeration
Discuss vaporized hydrogen peroxide sterilization parameters (temp,
concentration, time, other). - ANSWER-Temp: less than 122 degrees F
Concentration: 59%
Time: 28-55 minutes
Other: preset cycles
Discuss sterilization parameters for ethylene oxide (concentration, temp,
humidity, exposure time, aeration time) - ANSWER-Concentration: 450-1200
mg/l
Temp: 98-145 degrees F
Exposure time: 1-6 hours