Measurement of humidity can be accomplished - ANS ✓With a mirror whose
temperature can be controlled
A laboratory that uses an intrinsic standard may not simply use it without
verification, but must also - ANS ✓Participate in interlaboratory comparisons
or proficiency testing for that parameter
The relationship between the freezing point of water and the triple point of
water at standard atmospheric pressure is - ANS ✓The triple point is warmer
The SI unit used by torque wrenches is - ANS ✓N*m
What ratio of attenuator would be required to reduce a 1.0 mW signal to a
1.0 fW signal? - ANS ✓1 000 000 000 000 :1
The best reason that the local acceleration of gravity, little g, needs to be
known by a laboratory is because - ANS ✓It can affect many calibration results
When using saturated salt solutions to generate (source) humidity, it is
important to know the - ANS ✓Temperate
ASQ Calibration
, 2
ASQ
A client has submitted a 0 - 250 ft-lb, +- 5% of full scale, torque wrench to a
lab for calibration in the clockwise direction. The lab's 0 - 600 ft-lb, 0.25%
of reading, clockwise and counter-clockwise, electronic calibrator is out of
order. The lab also owns a spring-and-needle torque clockwise calibrator
that is working and is in calibration. However, it is 0 - 1000 ft-lb, 1% of
reading 2 ft-lb. Can this calibrator be used instead? - ANS ✓Yes, it is adequate
for this job
Which of the following parameters can be converted to relative humidity
using a table or calculation? - ANS ✓Dew point
ANSI/NCSL Z540.3 requires that calibration procedures shall contain
generic descriptions of the IM&TE and reference standards to be used so
that - ANS ✓The technician has all the required information to make a
substitution decision
A certificate of calibration issued by an NMI - ANS ✓May be considered
traceable without further evidence
Relative humidity is defined for which of the following gases or mixtures? -
ANS ✓Air
At what temperature do the Fahrenheit and Celsius scales indicate the same
numerical value? - ANS ✓-40 degrees
The derived SI unit, the Pascal, is a measure of - ANS ✓Pressure
10 dB is the same as - ANS ✓1.0 B
When a gas is in an enclosed box and the box gets smaller, the gas - ANS
✓Pressure increases
ASQ Calibration
, 3
ASQ
The sling on a psychrometer is used for - ANS ✓Swinging thermometers
around in the air
A laboratory puts together a toolbox with battery-operated measuring
devices to be used for calibration in remote locations. When on location, the
reference standard for the measurements being made - ANS ✓Is the best
device in the toolbox
A piston gage (deadweight tester) generates the most accurate pressure
values when it is - ANS ✓Level
deka - ANS ✓10^1 (da)
hecto - ANS ✓10^2 (h)
kilo - ANS ✓10^3 (k)
mega - ANS ✓10^6 (M)
giga - ANS ✓10^9 (G)
tera - ANS ✓10^12 (T)
peta - ANS ✓10^16 (P)
exa - ANS ✓10^18 (E)
zetta - ANS ✓10^21 (Z)