LH- Waveform Identification- ACNS Guidelines LATEST UPDATE
LH- Waveform Identification- ACNS Guidelines LATEST UPDATE Technical terms should be used in the impression selection of the EEG report. True or False - ANSWER False At the start of the study: -calibration is not longer required by the guidelines -10 seconds of bio-calibration should be performed -10 seconds of square wave calibration should be performed -10 seconds of square wave calibration and 10 seconds of bio-calibration should be performed - ANSWER 10 seconds of square wave calibraiton and 10 seconds of biocalibration should be performed In the 10 to 10 system, t6 is known as: -T8 -FP8 -T6 -P8 - ANSWER P8 A repeat ECI study can be done on an adult patient after: -24 hours -12 hours -4 hours -6 hours - ANSWER 6 hours The recommended band pass for an EEG is: -0.5-70Hz -0.5.80Hz -1-80Hz -1-70Hz - ANSWER 0.5-70Hz What items at minimum should be included on your record? (Check all that apply) -patient's age -patient's ID number -reason for the study -patient's date of birth -patient's name -technologist's name -date of the recording -name of the ordering physician - ANSWER Patient's age Patient's ID number Patient's name Technologist's name Date of the recording The maximum impedance for electrodes is: -2.5 kOhms -5 kOhms -15 kOhms -10 kOhms - ANSWER 10 kOhms