CCHT WATER| Questions with 100 % correct Answers | Verified
CHLORINE TESTING - REVERSE OSMOSIS (RO) MACHINES RUNS 15 MINUTES BEFORE PERFORMING ANY TOTAL CHLORINE TESTING 100 ML - THE CORRECT AMOUNT OF WATER TO COLLECT FOR TOTAL CHLORINE TESTING RPC ultra low total chlorine strips - STRIPS USED FOR CHLORINE TESTING HOW LONG ARE RPC STRIPS - IMMERSED IN SAMPLE WATER FOR 60 SECONDS WHEN SHOULD CHLORINE BE TESTED - PRIOR TO TREATMENTS IN THE MORNING AND EVERY 3-4 HOURS WHILE PATIENTS ARE DIALYZING RPC ULTRA -LOW TOTAL CHLORINE STRIPS TEST - CHLORINE IN THE WATER AFTER CARBON TANK AND POST RO 0.00 to 0.09 ppm - total chlorine levels that are patient safe AND DOCUMENT IN "TMS" OR PAPER LOG FIRST SAMPLE - SHOULD BE TAKEN FROM POST CARBON TANK 1 (WORKER ) SECOND SAMPLE - POST RO IF CHLORINE IS NOTED POST RO - DO NOT INITIATE ANY PATIENT TREATMENT, AND DOCUMENT PER FACILITY POLICY AND PROCEDURE CHLORINE BREAKTHROUGH - OCCURS POST CARBON TANK 1 (WORKER), IMMEDIATE ADDITIONAL TESTING SHOULD TAKE PLACE AFTER CARBON TANK 2 (POLISHER