Exam Preparation Newest 2026-2027 Real Exam With
Complete Questions And Correct Detailed
Answers| Just Released!!
Which of the following would not be safe to scan a patient
when
looking at it from the static magnetic field?
A. Bullet in the femur.
B. A piece of shrapnel in the mediastinum.
C. A BB in a patients sinus.
D. A small piece of metal in their abdominal rectus muscle.
- ANSWER-B.
A piece of shrapnel in the mediastinum.
Which of the following has the best chance for burning a
patient?
,A. BB
B. Pacing wire outside the RF field.
C. L 5 Fusion
D. Pacing wire inside the RF field. - ANSWER-D. Pacing
wire inside the RF field.
What is true of the time-varying gradient magnetic field
during a normal
MR scan?
,A. It always poses a risk to the safety of the patient.
B. It will cause a patient severe pain.
C. A spin echo sequence has the best chance to cause a
patient to have arrhythmia.
D. None of the choices. - ANSWER-D. None of the
choices.
,A pain pump implant manufacture instruction says the
following. The pump is located in the abdomen. Do not
scan directly over the implant.
Field strengths of 3T or less are safe. Limit heating to 2.2
W/kg or 0/7 deg. Celsius. You are scanning on a 3T
scanner and are performing a carotid MRA exam. Which
of he following precautions should you follow?
I. Do not give contrast media.
II. Stay in normal controlled operation mode.
III. Scan patient in a 1.5 T scanner.
IV. It is not safe to scan this patient.
A. I,II,III
B. IV
C. II
D. II, III - ANSWER-C. II
, Why is it important for the radiologist to have the final say
on if a patient with a questionable implant is safe to scan?
,A. They will be held responsible if harm is caused to the
patient if they are scanned.
B. They are qualified to make a benefit vs. risk
assessment, not the MR technologist.
C. They should be involved in the care of their patient.
D. All of the choices. - ANSWER-D. All of the choices.
Which of the following will produce the largest
susceptibility artifact?
A. Titanium clip
B. a cardiac stent.