Nuclear Medicine Instrumentation:
Exam 2 Questions and Answers
100% Pass
What is efficiency of counting/imaging? - CORRECT ANSWER:-How well
detector can... (goes dab, plus c)
1. geometry
2. operating voltage
3. energy resolution
4. scatter
5. dead time
6. absorption
7. background
plus...
collimation
, How do you test operating voltage for GM? - CORRECT ANSWER:-Battery
Check button (DC)
How do you test operating voltage for Cutie Pie? - CORRECT ANSWER:-Battery
Check button (DC)
What is operating voltage check called for Dx Calibrator? How do you check it? -
CORRECT ANSWER:-Constancy Test - button test with source
How do you test operating voltage for Gamma camera? Explain. - CORRECT
ANSWER:-Daily flood - checking that percentage is less than Uniformity
Correction Factor (%5)
What if UCF is over 5%? - CORRECT ANSWER:-redo flood and then call
biomed
What is the UCF value for the Gamma camera? - CORRECT ANSWER:-3-5 milli
counts
What is the operating voltage check called for the uptake probe? - CORRECT
ANSWER:-daily voltage check
What is the operating voltage check called for the well counter? - CORRECT
ANSWER:-daily voltage check
What units are used for GM? - CORRECT ANSWER:-counts/min or
disintegrations/min
What are the units of Cutie Pie? - CORRECT ANSWER:-millirems/hour
Exam 2 Questions and Answers
100% Pass
What is efficiency of counting/imaging? - CORRECT ANSWER:-How well
detector can... (goes dab, plus c)
1. geometry
2. operating voltage
3. energy resolution
4. scatter
5. dead time
6. absorption
7. background
plus...
collimation
, How do you test operating voltage for GM? - CORRECT ANSWER:-Battery
Check button (DC)
How do you test operating voltage for Cutie Pie? - CORRECT ANSWER:-Battery
Check button (DC)
What is operating voltage check called for Dx Calibrator? How do you check it? -
CORRECT ANSWER:-Constancy Test - button test with source
How do you test operating voltage for Gamma camera? Explain. - CORRECT
ANSWER:-Daily flood - checking that percentage is less than Uniformity
Correction Factor (%5)
What if UCF is over 5%? - CORRECT ANSWER:-redo flood and then call
biomed
What is the UCF value for the Gamma camera? - CORRECT ANSWER:-3-5 milli
counts
What is the operating voltage check called for the uptake probe? - CORRECT
ANSWER:-daily voltage check
What is the operating voltage check called for the well counter? - CORRECT
ANSWER:-daily voltage check
What units are used for GM? - CORRECT ANSWER:-counts/min or
disintegrations/min