Answers | Actual Complete Exam| Already Graded A+
1. Define: Un-restricted, semi-restricted Un-restricted = central control point,
and restricted access. mon- itors patients, personnel and
materials.
Street clothes are permitted, traflc not
lim- ited.
Semi = peripheral support areas of
surgical suite. Corridors leading to
restricted areas. Storage for clean and
sterile supplies. Traf- fic limited to
authorized personnel and pa- tients.
Personnel must wear surgical attire
and cover head and facial hair.
Restricted = OR'S, procedure rooms
and clean core area. surgical attire / hair
cover- ings required. Masks required
where open sterile supplies or
2. Air scrubbed persons are lo- cated.
Pressure:
OR- OR-
SPD- Positive
SPD-
Negative
3. Processing cycle in order.
Sterilize+Store+Deliver+Use+Return+Id
enti- fy+Clean+Assemble+
Package
4. Define: MSDS Material Safety Data Sheets
1/
36
,SPD Certification Prep with All Correct & 100% Verified
Answers | Actual Complete Exam| Already Graded A+
5. Temperatures: OR = 68* to 73* F (20* or 23*C)
OR Suite/General Work Decontamination = 60* to 65* F (16*
Areas Sterile Processing to 18* C)
2/
36
,SPD Certification Prep with All Correct & 100% Verified
Answers | Actual Complete Exam| Already Graded A+
Decontamination 2
6. Humidity: C
l
Peri-operative suite, instrumentation
a
and sterile areas
s
s
Sterile Storage areas
3
7. Air
Exchanges:
OR
Soilded decontamination, sterilizer
load- ing/un-loading area
Sterile storage/ Prep pack area
8. Why is point of use
decontamination im- portant?
9. Tap water should ideally have a
chloride concentration of ppm
or less.
10. Hands should be lathered and
scrubbed for at least?
11. Medical device classes:
Class 1
Class
3/
36
, SPD Certification Prep with All Correct & 100% Verified
Answers | Actual Complete Exam| Already Graded A+
Peri-operative suite = 30% to 60% Sterile storage =
Below 70%
OR = minimum 15 per hour/ reconmmend- ed 20 to 25
per hour
Decontamination/Sterilzer load-
ing/un-loading = minimum 10 per hour
Sterile storage/Prep pack = minimum 4 per hour
Starts the breakdown of Bio-burden process right away
100
15 seconds
Class 1 = Low risk devices. Most hand held surgical
instruments and ultrasonic clean- ers.
Class 2 = Pose potential risks. sterilization equipment, Sterility
indicators.
4/
36