Answers (2025-2026) Rated A+
A SSI PM is one that? - ANSWER-senses and pace in either the atrium or ventricle and
inhibits
All of the following functions are programmable in both VVI and DDD PM's, except? -
ANSWER-(X) AV interval duration
Sensitivity
Refractory
Output
An advantage of bipolar leads is? - ANSWER-They are less susceptible to EMI
As pulse duration is shortened below .3mse., stimulation threshold increases as
described by all of the following, except? - ANSWER-Energy
(x) Charge
Volts
Current
At the time of implant we want a threshold? - ANSWER-at .5ms to be less than 1 volt for
ventricle
Cardiac stimulation thresholds are expressed in terms of? - ANSWER-Voltage, current,
and energy.
Characteristic findings in patients with the PM syndrome include any of the following
except? - ANSWER-Febrile signs of PM's infection
Current PM's employ which of the following chemical batteries? - ANSWER-Lithium-
iodide
For a patient with evidence of sinus node dysfunction and intermittent heart block, which
of the following pacemakers would be inappropriate? - ANSWER-DDDR
(x) AAI
VVI
VVIR
for a wire fracture the resistance? - ANSWER-increases
For voltage of 5V, resistance of 500 ohms and pulse width of .5ms, calculate the energy
delivered? - ANSWER-25 microjoules
Hysterisis is? - ANSWER-The interval the PM waits after a sensed beat before pacing