CHLORINE TESTING - correct answers-REVERSE OSMOSIS (RO) MACHINES RUNS
15 MINUTES BEFORE PERFORMING ANY TOTAL CHLORINE TESTING
100 ML - correct answers-THE CORRECT AMOUNT OF WATER TO COLLECT FOR
TOTAL CHLORINE TESTING
RPC ultra low total chlorine strips - correct answers-STRIPS USED FOR CHLORINE
TESTING
HOW LONG ARE RPC STRIPS - correct answers-IMMERSED IN SAMPLE WATER
FOR 60 SECONDS
WHEN SHOULD CHLORINE BE TESTED - correct answers-PRIOR TO TREATMENTS
IN THE MORNING AND EVERY 3-4 HOURS WHILE PATIENTS ARE DIALYZING
RPC ULTRA -LOW TOTAL CHLORINE STRIPS TEST - correct answers-CHLORINE IN
THE WATER AFTER CARBON TANK AND POST RO
0.00 to 0.09 ppm - correct answers-total chlorine levels that are patient safe AND
DOCUMENT IN "TMS" OR PAPER LOG
FIRST SAMPLE - correct answers-SHOULD BE TAKEN FROM POST CARBON TANK
1 (WORKER )
SECOND SAMPLE - correct answers-POST RO
IF CHLORINE IS NOTED POST RO - correct answers-DO NOT INITIATE ANY
PATIENT TREATMENT, AND DOCUMENT PER FACILITY POLICY AND
PROCEDURE
CHLORINE BREAKTHROUGH - correct answers-OCCURS POST CARBON TANK 1
(WORKER), IMMEDIATE ADDITIONAL TESTING SHOULD TAKE PLACE AFTER
CARBON TANK 2 (POLISHER)
CARBON TANK 2 (POLISHER) - correct answers-IF TANK SHOWS RESULTS ABOVE
THE ACTION LEVEL IMMEDIATELY STOP TREATMENTS
BYPASS - correct answers-YOU WILL PUT PATIENTS IN THIS IF CARBON TANK 2
(POLISHER) TEST POSTIVE FOR CHLORINE , DOCUMENT TMS OR TCL- 2 LOG
EVERY 2 HOURS
MEDICAL DIRECTOR - correct answers-AFTER CHLORINE BREAKTHROUGH , WHO
MUST APPROVE TO RESUME NORMAL OPERATION AND SIGN INTO TMS OR
TCL -2 LOG
WHO TO NOTIFY IF CHLORINE BREAKTHROUGH - correct answers-CLINICAL
MANAGER, TECHNICAL SERVICES , MEDICAL DIRECTOR AND DIRECTOR OF
OPERATIONS
WATER HARDNESS - correct answers-WATER EXISTING THE WATER SOFTENR
PARTS PER MILLION (PPM) - correct answers-HOW WATER HARDNESS IS
MEASURED
20 PPM - correct answers-ACCEPTABLE LIMITS FOR WATER HARDNESS POST
SOFTNER
2026 QUESTIONS WITH ANSWERS
CHLORINE TESTING - correct answers-REVERSE OSMOSIS (RO) MACHINES RUNS
15 MINUTES BEFORE PERFORMING ANY TOTAL CHLORINE TESTING
100 ML - correct answers-THE CORRECT AMOUNT OF WATER TO COLLECT FOR
TOTAL CHLORINE TESTING
RPC ultra low total chlorine strips - correct answers-STRIPS USED FOR CHLORINE
TESTING
HOW LONG ARE RPC STRIPS - correct answers-IMMERSED IN SAMPLE WATER
FOR 60 SECONDS
WHEN SHOULD CHLORINE BE TESTED - correct answers-PRIOR TO TREATMENTS
IN THE MORNING AND EVERY 3-4 HOURS WHILE PATIENTS ARE DIALYZING
RPC ULTRA -LOW TOTAL CHLORINE STRIPS TEST - correct answers-CHLORINE IN
THE WATER AFTER CARBON TANK AND POST RO
0.00 to 0.09 ppm - correct answers-total chlorine levels that are patient safe AND
DOCUMENT IN "TMS" OR PAPER LOG
FIRST SAMPLE - correct answers-SHOULD BE TAKEN FROM POST CARBON TANK
1 (WORKER )
SECOND SAMPLE - correct answers-POST RO
IF CHLORINE IS NOTED POST RO - correct answers-DO NOT INITIATE ANY
PATIENT TREATMENT, AND DOCUMENT PER FACILITY POLICY AND
PROCEDURE
CHLORINE BREAKTHROUGH - correct answers-OCCURS POST CARBON TANK 1
(WORKER), IMMEDIATE ADDITIONAL TESTING SHOULD TAKE PLACE AFTER
CARBON TANK 2 (POLISHER)
CARBON TANK 2 (POLISHER) - correct answers-IF TANK SHOWS RESULTS ABOVE
THE ACTION LEVEL IMMEDIATELY STOP TREATMENTS
BYPASS - correct answers-YOU WILL PUT PATIENTS IN THIS IF CARBON TANK 2
(POLISHER) TEST POSTIVE FOR CHLORINE , DOCUMENT TMS OR TCL- 2 LOG
EVERY 2 HOURS
MEDICAL DIRECTOR - correct answers-AFTER CHLORINE BREAKTHROUGH , WHO
MUST APPROVE TO RESUME NORMAL OPERATION AND SIGN INTO TMS OR
TCL -2 LOG
WHO TO NOTIFY IF CHLORINE BREAKTHROUGH - correct answers-CLINICAL
MANAGER, TECHNICAL SERVICES , MEDICAL DIRECTOR AND DIRECTOR OF
OPERATIONS
WATER HARDNESS - correct answers-WATER EXISTING THE WATER SOFTENR
PARTS PER MILLION (PPM) - correct answers-HOW WATER HARDNESS IS
MEASURED
20 PPM - correct answers-ACCEPTABLE LIMITS FOR WATER HARDNESS POST
SOFTNER