EXAM 2025 QUESTIONS 100% SOLVED
Salinity is commonly measured using which unit? - ANSWERppt
How would you prepare 100 mL of 0.0400 M KCl from 0.200 M KCl stock solution? -
ANSWERWithdraw 20.0 mL of 0.200 M KCl, place it in a partially filled 100 mL
volumetric flask and add water until the final volume is 100 mL.
Which of the following is not considered Laboratory Personal Protective Equipment
(PPE)? - ANSWERHair Tie
Which of the following is a concern after a chemical spill? - ANSWEREnvironmental
release
Adequate engineering controls or PPE to protect staff during cleanup
Slip or Trip Hazard
Where should flammable liquids be stored? - ANSWERIn an approved flammable liquid
storage cabinet
What is an alternative way to indirectly measure the amount of dissolved ozone in an
aquarium? - ANSWERDPD test for chlorine or bromine
Which of the following disinfection methods has a low potential for creating disinfection
byproducts? - ANSWERUV
Which of the following do not contribute to Conductivity/Salinity? - ANSWEROxygen
What wavelength is considered part of the Visible Spectrum? - ANSWER500 nm
What is the volume of water in this graduated cylinder? - ANSWER6.6 mL
When analyzing ammonia, a technician analyzed spiked a sample containing 0.05 ppm
ammonia as N with a standard, adding 0.10 ppm ammonia as N. When she reads the
value of her LFM on the spectrophotometer, she sees a result of 0.05 ppm. Her LFB
was spiked with 0.10 ppm ammonia as N and her result was 0.09 ppm ammonia as N.
What corrective action is appropriate - ANSWERRe-prepare LFM and if it still fails,
consider using a different method
If a water quality laboratory technician records data incorrectly, how should they
respond? - ANSWERDraw a single line through the entry making sure that the
inaccurate information is still legible. Initial and date the entry. State the reason for the
error. Document the correct information.