RTBC- quality control Questions and Answers Latest Updated The central ray indicator must align to the actual center of the x-ray field with a tolerance of: 1% source-to-image distance (SID) The Source-to-Image Distance (SID) indicator misrepresent, the actual SID within a tolerance of: +- 2% of the SID The light field must correspond to the actual exposure field within a tolerance of: +- 2% of the SID " coincidence testing" of the x-ray to assembly is an evaluation of which of the following? Light field accuracy When the source to image distance (SID) is 80cm (31 inches), the actual century should be within what distance of the indicated x-ray field center? 0.8 cm (0.3 inches) At a source to image distance (SID) of 60 inches (152 cm), the central ray must align to the central ray indicator within: 0.6 inch (1.5cm) What is the primary purpose of regularly testing in calibrating the central ray alignment? Maintain the accuracy of the radiographic projection When the source the image distance (SID) is 80cm (31.5 inches), the actual SID should match the displayed SID within a range of: +-1.6 cm (0.6 inches) When the source the image distance (SID) is 80 cm (31.5 inches), the Light field should match the exposure field within a range of: +-1.6 cm (0.6 inches) When the source the image distance (SID) is 40 inches (102 cm), the actual SID should match the displayed SID within a range of: +- 0.8 inches (4.1 cm) Actual exposure time must be within what percentage of indicated exposure time? +-5% The actual exposure time must be accurate within what range when the assigned exposure time is 80 milliseconds (ms)? +-4.0 ms The actual exposure time must be accurate within what range when the assigned exposure time is 100 milliseconds (ms) +-5 ms The actual exposure time must be accurate within what range when the assigned exposure time is 72 milliseconds (ms) +-3.6 ms If the assigned exposure time is 60 milliseconds (ms), the actual exposure time should be no more than: 63.0 ms Which of these exposure times is outside acceptable limits when the exposure time on the operators consul, is set to 300 milliseconds (ms) 316 ms Consider this scenario: a medical physicist is testing an automatic exposure control (AEC) system. If an exposure through 1 cm of aluminum results in a receptor exposure of 10 micro gray, and exposure through 2 centimeters of aluminum should result in a receptor exposure of: 10 microgray (ugy) Consider this scenario: a medical physicist is testing an automatic exposure control (AEC) system. If an exposure through 10 cm of aluminum results in a receptor exposure of 100 micro gray, and exposure through 5 centimeters of aluminum should result in a receptor exposure of: 100 microgray (ugy) A radiograph is acquired using an automatic exposure controlled (AEC) unit with an exposure setting of 200 milliamperage (mA) and 60 Kilovoltage peak (kVp). If the initial receptor exposure is measured as 20 micro gray (Ugy), increasing the technique to 69 kVp will result in a receptor exposure of: 20 uGy A radiograph is acquired using an automatic exposure controlled (AEC) unit with an exposure setting of 200 milliamperage (mA) and 100 Kilovoltage peak (kVp). If the initial receptor exposure is measured as 50 micro gray (Ugy), increasing the technique to 300 kVp will result in a receptor exposure of: 50 Ugy the actual Kilovoltage peak (kVp) of an X-ray exposure must correspond to the selected kVp within a tolerance of what percentage? +-5% X-ray systems must demonstrate consistent exposure reproducibility within a tolerance of what percentage +-5% The X-ray tube output for identical technical parameters should not vary by more than what percentage? +-5% The milligray/milliampere-seconds (mGy/mAs) measurement between adjacent mA stations should not vary by more than: +-10% X-ray systems must demonstrate exposure linearity between adjacent milliamperage (mA) stations within a tolerance of what percentage? +-10% When selected tube is 80 Kilovoltage peak (kVp), the actual kVp must be accurate within a range of: +-4.0 kVp During annual control testing, the beam intensity measures 300 microgray (UGY) for an exposure performed at 100 Kilovoltage peak (kVp) and 10 milliampere-seconds (mAs). A second exposure performed with the same technique should match the first exposure within a tolerance of: +-15.0 uGy A diagnostic X-ray system must demonstrate exposure linearity within an error margin of what percentage? +-10% An exposure performed at 70 Kilovoltage peak (kVp) and 10 milliampere-seconds (mAs) produces a beam intensity of 342 microgray (Ugy). Calculate the Ugy/mAs. 34.2 uGy/mAs An X-ray system with unreliable exposure reproducibility may result in which two adverse effects? (Select two) -unnecessary patient dose -inadequate receptor exposure Three different detectors were expiases using a technique of 80 Kilovoltage peak (kVp) and 5 milliampere-seconds (mAs). The deviation index (DI) for each detector was recorded as: DI for detector 1= +2 DI for detector 2= +2 DI for detector 3= +1 the irregular DI for detector 3 indicates a quality control issue related to which of the following? Detector calibration Exposure of a small lead block is a technique used to evaluate which of the following? Reader erasure thoroughness A blank image receptor is exposed and processed. The resulting image shows an area on the plate with a lower radiation response. This may be a problem with which of the following? Plate uniformity Which aspect of an image receptor system is evaluated using a line-pairs pattern? Spatial resolution

Show more Read less
Institution
RTBC
Course
RTBC









Whoops! We can’t load your doc right now. Try again or contact support.

Written for

Institution
RTBC
Course
RTBC

Document information

Uploaded on
February 3, 2025
Number of pages
5
Written in
2024/2025
Type
Exam (elaborations)
Contains
Questions & answers

Subjects

Content preview

RTBC- quality control Questions and
Answers Latest Updated

The central ray indicator must align to the actual center of the x-ray field with a
tolerance of: - answer1% source-to-image distance (SID)

The Source-to-Image Distance (SID) indicator misrepresent, the actual SID within a
tolerance of: - answer+- 2% of the SID

The light field must correspond to the actual exposure field within a tolerance of: -
answer+- 2% of the SID

" Coincidence testing" of the x-ray to assembly is an evaluation of which of the
following? - answer Light field accuracy

When the source to image distance (SID) is 80cm (31 inches), the actual century should
be within what distance of the indicated x-ray field center? - answer0.8 cm (0.3 inches)

At a source to image distance (SID) of 60 inches (152 cm), the central ray must align to
the central ray indicator within: - answer0.6 inch (1.5cm)

What is the primary purpose of regularly testing in calibrating the central ray alignment?
- answerMaintain the accuracy of the radiographic projection

When the source the image distance (SID) is 80cm (31.5 inches), the actual SID should
match the displayed SID within a range of: - answer+-1.6 cm (0.6 inches)

When the source the image distance (SID) is 80 cm (31.5 inches), the Light field should
match the exposure field within a range of: - answer+-1.6 cm (0.6 inches)

When the source the image distance (SID) is 40 inches (102 cm), the actual SID should
match the displayed SID within a range of: - answer+- 0.8 inches (4.1 cm)

Actual exposure time must be within what percentage of indicated exposure time? -
answer+-5%

The actual exposure time must be accurate within what range when the assigned
exposure time is 80 milliseconds (ms)? - answer+-4.0 ms

The actual exposure time must be accurate within what range when the assigned
exposure time is 100 milliseconds (ms) - answer+-5 ms

, The actual exposure time must be accurate within what range when the assigned
exposure time is 72 milliseconds (ms) - answer+-3.6 ms

If the assigned exposure time is 60 milliseconds (ms), the actual exposure time should
be no more than: - answer63.0 ms

Which of these exposure times is outside acceptable limits when the exposure time on
the operators consul, is set to 300 milliseconds (ms) - answer316 ms

Consider this scenario: a medical physicist is testing an automatic exposure control
(AEC) system. If an exposure through 1 cm of aluminum results in a receptor exposure
of 10 micro gray, and exposure through 2 centimeters of aluminum should result in a
receptor exposure of: - answer10 microgray (ugy)

Consider this scenario: a medical physicist is testing an automatic exposure control
(AEC) system. If an exposure through 10 cm of aluminum results in a receptor exposure
of 100 micro gray, and exposure through 5 centimeters of aluminum should result in a
receptor exposure of: - answer100 microgray (ugy)

A radiograph is acquired using an automatic exposure controlled (AEC) unit with an
exposure setting of 200 milliamperage (mA) and 60 Kilovoltage peak (kVp). If the initial
receptor exposure is measured as 20 micro gray (Ugy), increasing the technique to 69
kVp will result in a receptor exposure of: - answer20 uGy

A radiograph is acquired using an automatic exposure controlled (AEC) unit with an
exposure setting of 200 milliamperage (mA) and 100 Kilovoltage peak (kVp). If the initial
receptor exposure is measured as 50 micro gray (Ugy), increasing the technique to 300
kVp will result in a receptor exposure of: - answer50 Ugy

the actual Kilovoltage peak (kVp) of an X-ray exposure must correspond to the selected
kVp within a tolerance of what percentage? - answer+-5%

X-ray systems must demonstrate consistent exposure reproducibility within a tolerance
of what percentage - answer+-5%

The X-ray tube output for identical technical parameters should not vary by more than
what percentage? - answer+-5%

The milligray/milliampere-seconds (mGy/mAs) measurement between adjacent mA
stations should not vary by more than: - answer+-10%

X-ray systems must demonstrate exposure linearity between adjacent milliamperage
(mA) stations within a tolerance of what percentage? - answer+-10%

Get to know the seller

Seller avatar
Reputation scores are based on the amount of documents a seller has sold for a fee and the reviews they have received for those documents. There are three levels: Bronze, Silver and Gold. The better the reputation, the more your can rely on the quality of the sellers work.
julianah420 Phoenix University
View profile
Follow You need to be logged in order to follow users or courses
Sold
655
Member since
2 year
Number of followers
324
Documents
33814
Last sold
1 day ago
NURSING,TESTBANKS,ASSIGNMENT,AQA AND ALL REVISION MATERIALS

On this page, you find all documents, package deals, and flashcards offered by seller julianah420

4.3

149 reviews

5
101
4
20
3
8
2
5
1
15

Recently viewed by you

Why students choose Stuvia

Created by fellow students, verified by reviews

Quality you can trust: written by students who passed their tests and reviewed by others who've used these notes.

Didn't get what you expected? Choose another document

No worries! You can instantly pick a different document that better fits what you're looking for.

Pay as you like, start learning right away

No subscription, no commitments. Pay the way you're used to via credit card and download your PDF document instantly.

Student with book image

“Bought, downloaded, and aced it. It really can be that simple.”

Alisha Student

Frequently asked questions