0.1-10 uV - Electrical amplitude of SSEPs is A. 10-100 uV B. 0.1 to 10 uV C. 1000 to 10,000 uV D. 100 to 1000 uV 0.1-5000 Hz - What is the minimum required bandpass for an EP system 0.2ms - The duration of a monophasic, rectangular pulse used for somatosensory stimulation should be: A. 0.05 ms B. 0.2 ms C. 0.5 ms D. 1 ms 10 mS - What is the minimum analysis time for ABRs? 10 msec - What is the minimum acceptable sweep for ABRs? 10 uA - What is the maximum allowable leakage current from the isolated amplifier inputs through the patient leads? 100-1000 V - A typical stimulus intensity required to obtain TCMEPs is A. 0.1-1 V B. 10-100 V C. 1-100 V