Principles Of DNP Anesthesia Practice Questions And Answers 100% Solved
Principles Of DNP Anesthesia Practice Questions And Answers 100% Solved After the anesthesia machine receives and mixes gases, it sends it to the vaporizer where it is mixed with anesthetics and then delivered where?️️Common gas outlet (CGO) The anesthesia machine gas pathways are divided up into what three systems?️️Highpressure system: Includes parts upstream of the first stage regulator; gas pressures are between 45 and 2200 psig Intermediate pressure system: Includes parts DOWNSTREAM of the first stage regulator and UPSTREAM the gas flow control valves; gas pressures are between 16 and 55 psig Low pressure system: Includes all parts downstream of the gas flow control valves where pressures are normally slightly greater than atmospheric pressure What is the diameter index safety system (DISS)?️️The pipeline inlets into the anesthesia machine are gas specific by a national standard (ensures that pipeline gas can not be connected to the wrong spot) What does the DISS system specify?️️The internal bore dimensions of the anesthesia machine inlet which only compatible hose connectors with matching outer dimensions can insert, oxygen has a different design than all others