What is the DESIRABLE sampling rate for EEG, EOG, ECG, & snoring
sounds? - ✔ ✔ 500Hz
What is the MINIMUM sampling rate for EEG, EOG, ECG, & snoring
sounds? - ✔ ✔ 200Hz
What is the DESIRABLE sampling rate for airflow, nasal pressure, end-
tidal PCO2, PAP device flow, rib cage and abdominal movements? - ✔
✔ 100Hz
What is the MINIMUM sampling rate for airflow, nasal pressure, end-
tidal PCO2, PAP device flow, rib cage and abdominal movements? - ✔
✔ 25Hz
What is the DESIRABLE sampling rate for oximetry or transcutaneous
PCO2? - ✔ ✔ 25hz
What is the MINIMUM sampling rate for oximetry or transcutaneous
PCO2? - ✔ ✔ 10Hz
What is the DESIRABLE and MINIMUM sampling rate for body position?
- ✔ ✔ 1Hz
What is the recommended LOW frequency filter for EEG & EOG? - ✔
✔ 0.3Hz
What is the recommended HIGH frequency filter for EEG & EOG? - ✔
✔ 35Hz
, What is the recommended LOW frequency filter setting for EMG? - ✔
✔ 10Hz
What is the recommended HIGH frequency filter setting for EMG? - ✔
✔ 100Hz
What is the recommended LOW frequency filter setting for oronasal
thermal flow & thoracoabdominal belt signals? - ✔ ✔ 0.1 Hz
What is the recommended HIGH frequency filter setting for oronasal
thermal flow & thoracoabdominal belt signals? - ✔ ✔ 15Hz
What is the recommended LOW frequency filter setting for nasal
pressure? - ✔ ✔ DC or less than or equal to 0.03 Hz
What is the recommended HIGH frequency filter setting for NASAL
pressure? - ✔ ✔ 100Hz
What is the recommended LOW frequency filter setting for PAP device
flow? - ✔ ✔ DC
What is the recommended HIGH frequency filter setting for PAP device
flow? - ✔ ✔ DC
What is the recommended LOW frequency filter setting for Snoring? -
✔ ✔ 10Hz
What is the recommended HIGH frequency filter setting for snoring? -
✔ ✔ 100Hz
What is the preferred impedence level? - ✔ ✔ 5 ohms or less and
relatively equal
What impedence level is acceptable for limb EMG? - ✔ ✔ 10 ohms