100% de satisfacción garantizada Inmediatamente disponible después del pago Tanto en línea como en PDF No estas atado a nada 4,6 TrustPilot
logo-home
Examen

CH. 29 PREVENTING AND SOLVING ANESTHESIA MACHINE PROBLEMS

Puntuación
-
Vendido
-
Páginas
8
Grado
A+
Subido en
22-08-2025
Escrito en
2025/2026

CH. 29 PREVENTING AND SOLVING ANESTHESIA MACHINE PROBLEMS

Institución
ANESTHESIA MACHINE
Grado
ANESTHESIA MACHINE









Ups! No podemos cargar tu documento ahora. Inténtalo de nuevo o contacta con soporte.

Escuela, estudio y materia

Institución
ANESTHESIA MACHINE
Grado
ANESTHESIA MACHINE

Información del documento

Subido en
22 de agosto de 2025
Número de páginas
8
Escrito en
2025/2026
Tipo
Examen
Contiene
Preguntas y respuestas

Temas

Vista previa del contenido

CH. 29 PREVENTING AND SOLVING ANESTHESIA
MACHINE PROBLEMS


_____ and _____ have collaborated to develop recommendations for checking out an
anesthesia machine prior to administering an anesthetic - Answer -The American
Society of Anesthesiologists (ASA) and the Food and Drug Administration (FDA)

An anesthesia machine's main function will always be to provide - Answer -a controlled
supply of oxygen and other anesthetic gases to the patient during surgery

The high-pressure circuit starts where - Answer -the gas enters the machine and ends
at the flow control valves

a full tank of N2O has ___ psi - Answer -745 psi

O2 tanks have ___ psi when full - Answer -2,200

Fail-safe valve it serves to - Answer -decrease the supply of nitrous oxide if the oxygen
pressure drops

The low-pressure system of the anesthesia machine begins - Answer -downstream of
the flow valves

If _______ in the anesthesia machine is detected, the following items should be
examined:
Check for low pressure leak.
Check for a circuit disconnect starting with the patient and working back to the machine.
Inspect bellows.
Inspect CO2 canister.
Check lids/coupling of vaporizers.
Inspect the flow sensor.
Check oxygen pressure from pipeline/cylinder.
Check for incompetent ventilator relief valve. - Answer -low pressure

The most common cause for a low-pressure alarm is a - Answer -leak in the low-
pressure circuit

The most common cause of a leak is a - Answer -disconnection or partial disconnection
in the patient circuit

High negative pressure in the scavenging system can also cause ____ pressure in the
breathing circuit - Answer -high

, An incompetent ventilator relief valve can lead to _____ because gas is delivered to the
scavenging system instead of to the patient when the bellows collapse - Answer -
hypoventilation

A common cause for a hypoxic gas mixture alarm is _______ of the oxygen sensor at
the time of machine checkout - Answer -improper calibration

These steps are to fix machine if:
Turn on 100% oxygen and check for proper reading from the oxygen sensor.
Recalibrate the O2 sensor.
Replace O2 electrode if it does not appropriately recalibrate.
Check for main pipeline crossover. - Answer -Hypoxic Gas Mixture

These steps are to fix machine if:
Check for a leak.
Check for a stuck bobbin - Answer -Flowmeter Problems

These steps are to fix machine if:
Check for low-pressure leak.
Check for a circuit disconnect starting with the patient and working back to the machine.
Inspect bellows.
Inspect CO2 canister.
Check lids/coupling of vaporizers.
Inspect the flow sensor.
Inspect gas analyzer sampling rate.
Check oxygen pressure from pipeline/cylinder.
Check for an incompetent ventilator relief valve.
Rule out ventilator problems by switching to manual ventilation mode.
Rule out anesthesia machine problems by switching to an Ambu bag and/or an
alternative form of anesthetic delivery (e.g., intravenous anesthesia). Perform a leak
test. - Answer -Ventilator will not fill: The most common cause is a leak in the system.
Perform the same steps as above to investigate a low-pressure alarm. In addition,
check for an incompetent ventilator relief valve.

These steps are to fix machine if:
Look for inspiratory reverse flow/negative inspiratory flow alarm.
Check for high airway pressures.
Check for an incompetent expiratory valve. - Answer -Ventilator will not deliver a breath

These steps are to fix machine if:
Check the breathing circuit from the machine to the patient for kinking or any type of
obstruction
Check the ETT for obstruction (kinking, secretions, mainstem intubation, etc.)
Consider patient pathophysiology (bronchospasm, tension pneumothorax, stone chest,
etc.)
Check pop-off valve
$14.49
Accede al documento completo:

100% de satisfacción garantizada
Inmediatamente disponible después del pago
Tanto en línea como en PDF
No estas atado a nada


Documento también disponible en un lote

Conoce al vendedor

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
GEEKA YALA UNIVERSITY
Seguir Necesitas iniciar sesión para seguir a otros usuarios o asignaturas
Vendido
2009
Miembro desde
3 año
Número de seguidores
1447
Documentos
48318
Última venta
18 horas hace

3.8

344 reseñas

5
171
4
61
3
44
2
16
1
52

Recientemente visto por ti

Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes