RTBC- Quality Control: Questions With Fully
Explained Solutions
The central ray indicator must align to the actual center of the x-ray
field with a tolerance of: Correct Answer - 1% source-to-image
distance (SID)
The Source-to-Image Distance (SID) indicator misrepresent, the
actual SID within a tolerance of: Correct Answer - +- 2% of the
SID
The light field must correspond to the actual exposure field within a
tolerance of: Correct Answer - +- 2% of the SID
" coincidence testing" of the x-ray to assembly is an evaluation of
which of the following? Correct Answer - Light field accuracy
When the source to image distance (SID) is 80cm (31 inches), the
actual century should be within what distance of the indicated x-ray
field center? Correct Answer - 0.8 cm (0.3 inches)
At a source to image distance (SID) of 60 inches (152 cm), the
central ray must align to the central ray indicator within: Correct
Answer - 0.6 inch (1.5cm)
What is the primary purpose of regularly testing in calibrating the
central ray alignment? Correct Answer - Maintain the accuracy
of the radiographic projection
When the source the image distance (SID) is 80cm (31.5 inches), the
actual SID should match the displayed SID within a range of:
Correct Answer - +-1.6 cm (0.6 inches)
, When the source the image distance (SID) is 80 cm (31.5 inches), the
Light field should match the exposure field within a range of:
Correct Answer - +-1.6 cm (0.6 inches)
When the source the image distance (SID) is 40 inches (102 cm), the
actual SID should match the displayed SID within a range of:
Correct Answer - +- 0.8 inches (4.1 cm)
Actual exposure time must be within what percentage of indicated
exposure time? Correct Answer - +-5%
The actual exposure time must be accurate within what range when
the assigned exposure time is 80 milliseconds (ms)? Correct
Answer - +-4.0 ms
The actual exposure time must be accurate within what range when
the assigned exposure time is 100 milliseconds (ms) Correct
Answer - +-5 ms
The actual exposure time must be accurate within what range when
the assigned exposure time is 72 milliseconds (ms) Correct
Answer - +-3.6 ms
If the assigned exposure time is 60 milliseconds (ms), the actual
exposure time should be no more than: Correct Answer - 63.0
ms
Which of these exposure times is outside acceptable limits when the
exposure time on the operators consul, is set to 300 milliseconds
(ms) Correct Answer - 316 ms
Consider this scenario: a medical physicist is testing an automatic
exposure control (AEC) system. If an exposure through 1 cm of
aluminum results in a receptor exposure of 10 micro gray, and
Explained Solutions
The central ray indicator must align to the actual center of the x-ray
field with a tolerance of: Correct Answer - 1% source-to-image
distance (SID)
The Source-to-Image Distance (SID) indicator misrepresent, the
actual SID within a tolerance of: Correct Answer - +- 2% of the
SID
The light field must correspond to the actual exposure field within a
tolerance of: Correct Answer - +- 2% of the SID
" coincidence testing" of the x-ray to assembly is an evaluation of
which of the following? Correct Answer - Light field accuracy
When the source to image distance (SID) is 80cm (31 inches), the
actual century should be within what distance of the indicated x-ray
field center? Correct Answer - 0.8 cm (0.3 inches)
At a source to image distance (SID) of 60 inches (152 cm), the
central ray must align to the central ray indicator within: Correct
Answer - 0.6 inch (1.5cm)
What is the primary purpose of regularly testing in calibrating the
central ray alignment? Correct Answer - Maintain the accuracy
of the radiographic projection
When the source the image distance (SID) is 80cm (31.5 inches), the
actual SID should match the displayed SID within a range of:
Correct Answer - +-1.6 cm (0.6 inches)
, When the source the image distance (SID) is 80 cm (31.5 inches), the
Light field should match the exposure field within a range of:
Correct Answer - +-1.6 cm (0.6 inches)
When the source the image distance (SID) is 40 inches (102 cm), the
actual SID should match the displayed SID within a range of:
Correct Answer - +- 0.8 inches (4.1 cm)
Actual exposure time must be within what percentage of indicated
exposure time? Correct Answer - +-5%
The actual exposure time must be accurate within what range when
the assigned exposure time is 80 milliseconds (ms)? Correct
Answer - +-4.0 ms
The actual exposure time must be accurate within what range when
the assigned exposure time is 100 milliseconds (ms) Correct
Answer - +-5 ms
The actual exposure time must be accurate within what range when
the assigned exposure time is 72 milliseconds (ms) Correct
Answer - +-3.6 ms
If the assigned exposure time is 60 milliseconds (ms), the actual
exposure time should be no more than: Correct Answer - 63.0
ms
Which of these exposure times is outside acceptable limits when the
exposure time on the operators consul, is set to 300 milliseconds
(ms) Correct Answer - 316 ms
Consider this scenario: a medical physicist is testing an automatic
exposure control (AEC) system. If an exposure through 1 cm of
aluminum results in a receptor exposure of 10 micro gray, and