Mild sensory effects due to static magnetic fields include: - (answer)• Vertigo

• Nausea

• Taste Sensations



Sensory effects of the static magnetic field increase or decrease with field strength? -

(answer)Sensory effects are related to the strength of the magnetic field with few symptoms

at 1.5T and more appearing approaching a 4T field strength

眼睛是人类最重要

Explain what t-wave artefact is and why it occurs - (answer)Any charged particle moving

through a magnetic field will have a current induced via this interaction. As blood, a

conductive fluid flows through a magnetic field it induces an electrical biopotential. This

electrical biopotential can be seen on the ECG waveform of patients in MRI. It is

demonstrated by an increase in the T-wave amplitude on ECG. This change in T-wave

amplitude is directly proportional to the strength of the Static Filed. It is important that MR

radiographers are aware of this as elevated T-waves are also associated with ischemia and

myocardial infarction.



What risks does the static magnetic field pose in terms of implants and objects? -

(answer)Static Fields also pose hazards through the displacement of ferro-magnetic

implants and objects. The static field imparts both a translation (attractive force) and a

,torque (twisting force). The strong nature of the Static Field also has the potential to disrupt

the function of Cardiac Pacemakers. Anyone entering beyond the 5G line should be

screened for contraindications to strong Magnetic Fields. Also be aware of what is being

taken over the 5G line.



What strategies can be employed to reduce the risks of the static magnetic field impacting

on implants and objects? - (answer)• Ferromagnetic objects may become airborne - 40

km/hr terminal velocity possible at 1.5T

• Test any metal objects with a hand held magnet before allowing them to enter the room.

• Keep the general public behind the 5G line. •

Possible contra-indications include pacemakers, aneurysm clips, intra-vascular coils and
眼睛是人类最重要
stents, heart valves, penile implants, cochlear implants, ocular implants (Fatio eyelid

springs), neuro stimulators, bone growth stimulators, drug infusion pumps.



What effect do gradient magnetic fields have on patient safety? - (answer)Faraday's Law

dictates that changing magnetic fields will generate a voltage and a current in a conductor.

The switching of the gradients induces electrical currents in conductive tissues. Voltage

generation will occur during the rise and fall times - ie. while the field is changing. The

largest voltages will be generated at the periphery where the gradient amplitudes are

highest. The most common result of these voltages is Peripheral Nerve Stimulation (PNS).

Other side-effects are listed below;

• magnetophosphenes - electromagnetically induced flashes of light

• seizures

,• tissue heating

• peripheral nerve stimulation

• muscle contractions

• cardiac arrhythmias

• some degree of patient heating has been shown to be a result of Gradient Magnetic effects.

However, this has been shown to be so low as to be negligible



What strategies can be employed to keep gradient effects to a minimum? - (answer)In order

to keep these effects to a minimum the following is recommended.

• keep dB/dT less than those required to produce peripheral nerve stimulation.

• max limit for dB/dT of 6 Tesla / sec
眼睛是人类最重要


What is the cause of acoustic noise during an MRI scan? - (answer)The characteristic noise

that is heard during an MRI scan is the result of the interaction of Lorentz forces generated

by the Gradient Coil when a current is pulsed through them in the presence of a Static

Magnetic Field. The noise results as these forces are so strong as to twist the coils on their

mounts. In some cases this can be in excess of 100dB particularly when running EPI

sequences. Thus all patients need to be provided hearing protection.



What is the greatest concern in terms of MR safety? - (answer)RF effects are of the greatest

concern in terms of MR safety. The main effect is the deposition of energy resulting in tissue

heating. Of particular concern are heat sensitive organs such as the eyes and testes. One of

the major concerns is where a patient has a metallic implant which results in greater heating.

, RF field safety concerns increase or decrease with increasing field strength> - (answer)The

amount of energy absorbed will increase with frequency of the RF field delivered and is

therefore greater at higher field strengths



What is SAR? - (answer)SAR= Specific Absorption Rate - the term used to describe energy

dissipation



SAR is expressed in: - (answer)Watts / kg



SAR will depend on factors including: - (answer)• frequency (& therefore the field strength)
眼睛是人类最重要
• RF pulse - where power deposited is proportional to (flip angle / 90)

• TR

• RF coil - transmit / receive or receive only

• volume of tissue within the coil

• conductivity of the tissue



What determines the SAR calculation? - (answer)The SAR is calculated for each sequence

on the basis of the sequence parameters and the patient weight



What are the FDA recommendations regarding SAR? - (answer)• Maximum SAR of 0.4

W/kg (whole body)

• 3.2 W/kg (head)

Información del documento

Subido en
5 de diciembre de 2024
Número de páginas
76
Escrito en
2024/2025
Tipo
Examen
Contiene
Preguntas y respuestas
$18.49

¿Documento equivocado? Cámbialo gratis Dentro de los 14 días posteriores a la compra y antes de descargarlo, puedes elegir otro documento. Puedes gastar el importe de nuevo.
Escrito por estudiantes que aprobaron
Inmediatamente disponible después del pago
Leer en línea o como PDF

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
waynev
4.9
(241)
Vendido
2682
Seguidores
5
Artículos
1288
Última venta
2 meses hace

Reseñas de compradores verificados




Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes