CNOR 2025 Zander Notes Name the 3 Spaulding Classification Systems Critical Semi-Critical Non-Critical Meaning of Critical Spaulding Classification System Items MUST be sterile. Items will enter -sterile tissue- or -vascular system- Examples of items in Critical Spaulding Classification System -Instruments- -Laparoscopes- -Arthroscopes- -*Cutting endoscope accessories for biopsy*- -Cardiac Caths- -IV caths- - Needles- -*Urinary Caths- Meaning of Semi-Critical Spaulding Classification System Items SHOULD be sterile. Items will contact -broken skin- or -MM- If the manufacturers instructions say it WON'T tolerate sterilization, they CAN be HLD (but only IF MF SAYS). -Spores survive HLD- Examples of items in Semi-Critical Spaulding Classification System -Bronchoscopes -GI Scopes/GI endoscopes -RT equipment (laryngoscope blades) -Anesthesia equipment Meaning of Non Critical Spaulding Classification System -Will be in contact with INTACT SKIN -Intermediate or Low Level Disinfection or Cleaning. -Disinfect with an FDA approved disinfectant that's effective against HEP B. Examples of items in Non Critical Spaulding Classification System -Furniture -Laundry/Linens -OR bed -BP Cuff/Leads/Monitors/Stethoscopes -Tourniquets Requirement for decontamination area: -Must have a physical WALL or SEPARATION between decontam and processing areas. -This includes Immediate Use Areas. -Why: Because decontam requires NEGATIVE PRESSURE Sterile Processing Workflow Decontamination (always 1st) -- Prep&Pack --Sterilization --Distribute/Store Most important step to prevent infection: Cleaning and Pre-treating with a non-drying enzymatic solution in order to avoid biofilm, which encapsulates spores. Decontamination processes -Disassemble everything -Brush lumens/channels using a brush based on mfr instruction for correct diameter and length of lumen -Brush lumens/channels/crevices/joints UNDER WATER -Automated (washers/ultrasonic) more effective than manual cleaning. Two Types of Steam Sterilization -Pre-Vac Autoclave/Dynamic Air Removal -Gravity Displacement PreVac settings 270-276 3 min (NON porous instruments/regular instruments) 4 min (POROUS or LUMEN instruments, heavy plastic, rubber, wooden handles, poole suction) PreVac/Dynamic Air Removal setting for porous (heavy plastic, rubber, wood handels) or lumen instruments (poole suction/channels) 270-276 for 4 min PreVac/Dynamic Air Removal setting for Non Porous instruments 270-276 for 3 min How does PreVac/Dynamic Air Removal autoclave work? -Sucks air out of chamber with a VACCUM then steam is released and snaps down onto inst/trays. -DRAINS MUST BE CLEAN -Most common. Most effective. Gravity Displacement Autoclave settings 270-276

Vista previa del contenido

CNOR 2025 Zander Notes

Name the 3 Spaulding Classification Systems ✅Critical

Semi-Critical

Non-Critical



Meaning of Critical Spaulding Classification System ✅Items MUST be sterile.

Items will enter -sterile tissue- or -vascular system-



Examples of items in Critical Spaulding Classification System ✅-Instruments-

-Laparoscopes- -Arthroscopes-

-*Cutting endoscope accessories for biopsy*-

-Cardiac Caths-

-IV caths- - Needles-

-*Urinary Caths-



Meaning of Semi-Critical Spaulding Classification System ✅Items SHOULD be sterile.

Items will contact -broken skin- or -MM-



If the manufacturers instructions say it WON'T tolerate sterilization, they CAN be HLD (but only IF MF
SAYS).

-Spores survive HLD-



Examples of items in Semi-Critical Spaulding Classification System ✅-Bronchoscopes

-GI Scopes/GI endoscopes

-RT equipment (laryngoscope blades)

-Anesthesia equipment

,Meaning of Non Critical Spaulding Classification System ✅-Will be in contact with INTACT SKIN

-Intermediate or Low Level Disinfection or Cleaning.

-Disinfect with an FDA approved disinfectant that's effective against HEP B.



Examples of items in Non Critical Spaulding Classification System ✅-Furniture

-Laundry/Linens

-OR bed

-BP Cuff/Leads/Monitors/Stethoscopes

-Tourniquets



Requirement for decontamination area: ✅-Must have a physical WALL or SEPARATION between
decontam and processing areas.

-This includes Immediate Use Areas.

-Why: Because decontam requires NEGATIVE PRESSURE



Sterile Processing Workflow ✅Decontamination (always 1st) --> Prep&Pack -->Sterilization

-->Distribute/Store



Most important step to prevent infection: ✅Cleaning and Pre-treating with a non-drying enzymatic
solution in order to avoid biofilm, which encapsulates spores.



Decontamination processes ✅-Disassemble everything

-Brush lumens/channels using a brush based on mfr instruction for correct diameter and length of
lumen

-Brush lumens/channels/crevices/joints UNDER WATER

-Automated (washers/ultrasonic) more effective than manual cleaning.



Two Types of Steam Sterilization ✅-Pre-Vac Autoclave/Dynamic Air Removal

-Gravity Displacement

,PreVac settings ✅270-276



3 min (NON porous instruments/regular instruments)



4 min (POROUS or LUMEN instruments, heavy plastic, rubber, wooden handles, poole suction)



PreVac/Dynamic Air Removal setting for porous (heavy plastic, rubber, wood handels) or lumen
instruments (poole suction/channels) ✅270-276 for 4 min



PreVac/Dynamic Air Removal setting for Non Porous instruments ✅270-276 for 3 min



How does PreVac/Dynamic Air Removal autoclave work? ✅-Sucks air out of chamber with a VACCUM
then steam is released and snaps down onto inst/trays.

-DRAINS MUST BE CLEAN

-Most common. Most effective.



Gravity Displacement Autoclave settings ✅270-276



3 min: NonPorous/reg instruments



10 min: Porous/Lumens



Gravity Displacement autoclave settings for porous/lumens ✅270-276 10 min



Gravity Displacement autoclave settings for NonPorous/reg inst ✅270-276 3 min



How does Gravity Displacement autoclave work? ✅Steam pushes air down to a thermometer in the
drain, the air exits the chamber, drain closes, cycle starts.

, -Not great for trays or packages.



-Rarely used.



Both PreVac/Dynamic Air Removal and Gravity time and temperature requirements are for the
instrument and autoclave only.

For certain containers/systems such as One Tray system or containers with lids: ✅I will follow the
manufacturers instructions!



3 ways to avoid condensation with steam sterilization items: ✅- Allow airflow between trays and pkgs
and all the way around.

-Remove loads immediately (don't crack the door to cool).

-Don't place items on cooling racks or solid surfaces (can place on lined surfaces).



How many classes of Chemical Integrators? ✅6



What kind of chemical integrators react to heat? ✅Class 1 Chemical Indicator

-Reacts to HEAT only. Verifies exposure only.

-Tape-Locks-Filter dots



What kind of chemical integrator is an External Indicator? ✅Class 1 Chemical Indicator

External Indicator that reacts to Heat only

-Tape-Locks-Filter dots



What class of Chemical Indicator is the Bowie Dick, and what does it react to? ✅-Class 2 Chemical
Indicator

-Reacts to PRESSURE and ONLY indicates a VACUUM was created.

-Used only with PreVac/Dynamic Air Removal cycles

Información del documento

Subido en
10 de octubre de 2024
Número de páginas
90
Escrito en
2024/2025
Tipo
Examen
Contiene
Preguntas y respuestas
$10.99

¿Documento equivocado? Cámbialo gratis Dentro de los 14 días posteriores a la compra y antes de descargarlo, puedes elegir otro documento. Puedes gastar el importe de nuevo.
Escrito por estudiantes que aprobaron
Inmediatamente disponible después del pago
Leer en línea o como PDF

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
CertifiedGrades
3.9
(38)
Vendido
149
Seguidores
61
Artículos
8729
Última venta
1 mes hace




Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes