CRCST Certification: Chapter 14 Quiz
When loading a steam sterilizer, basins should be placed ____________. - Answer- on edge The higher the bioburden on an object the ________________. - Answer- the more difficult it will be to sterilize True or False? When combining loads, hard goods should be placed on the top shelves to allow for more efficient removal of the condensate. - Answer- False True or False? The steam sterilization process can be affected by the design of the medical device being sterilized. - Answer- True Central Service technicians need to understand the anatomy of a steam sterilizer to ____________________. - Answer- understand how the sterilizer operates The weakest part of a steam sterilizer is the __________. - Answer- door Three of the main phases of a terminal steam sterilizer cycle are _____________, ____________ and _______________. - Answer- conditioning, exposure and exhaust The most common reason for steam sterilization failure is ______________. - Answer- lack of steam contact with the instrument The coolest place in a steam sterilizer is the _____________. - Answer- thermostatic trap True or False? Steam flush pressure pulse sterilizers are a type of gravity sterilizers. - Answer- False Factors that can cause sterilant contact failure with the instrument are ________________ and a _______________. - Answer- crowded loads / clogged drain strainer One of the most frequent causes of a clogged drain screen is ___________. - Answer- tape True or Fal