●● What does increasing the compression ratio do, and when would you
use this?
Answer: Soft sound gain is unaltered, but gain for moderate/loud sounds
is lowered. Most often used if a PT complains loud/moderate sounds are
too loud or background noise is too loud. ** sometimes you use this on
just certain frequencies, for example, high frequencies if they complain
dishes are too loud, or low frequencies if they complain their voice is
echo-y


●● What happens when you raise a knee point, and when would you do
this?
Answer: You will decrease gain for soft sounds and raise gain for
moderate/loud sounds. Most often used if they hear soft sounds better
than moderate ones ( like hearing the table across from them better than
their own in a restaurant) or when they complain about hearing a HA's
circuit noise.


●● What does lowering the kneepoint do, and when would you do this?
Answer: Lowering the KP will lower gain for moderate and loud sounds,
and increase gain for soft sounds below the original KP. Mostly used
when a PT complains they can't hear soft sounds like soft voices, bird
chirping, or ticking clocks.

,●● What are the two types of compression?
Answer: Input compression - when the input level of a HA determines
when compression is activated.
Output compression - when the output (input + gain) of the HA
determines when compression is activated.


●● What is Wide Dynamic Range Compression?
Answer: (WDRC) This is what most HA's utilize today, and allows the
signal to be packaged withing a user's DR. Typically WDRC HA's have
low (35-55 dB SPL) KPs and low (1:1 - 1:4) ratios


●● What is the Compression Ratio?
Answer: The ratio of change between the input and output of a HA. This
determines how much compression will occur.


●● What is the Kneepoint? (KP)
Answer: The lowest level at which compression is applied


●● What is Harmonic Distortion?
Answer: Harmonic distortion occurs when a single frequency is
presented to the input of a hearing aid and the output contains the
original frequency plus additional undesired frequencies that are
harmonically related to the original frequency.

,ANSI uses a harmonic distortion test that measures the distortion rate at
specific Hz.


●● Frequency Response Curve
Answer: Used in ANSI testing. A gain curve is run with the input sound
at 60 dB SPL to simulate conversation level. You get the frequency
range calculation from this test.


●● Reference Test Gain (RTG)
Answer: Used in ANSI testing. Input sound is set at 60 dB SPL to
simulate conversation level; and then the HFA gain is adjusted to HFA-
OSPL90 -77 dB. The answer is the RTG.


●● Full On Gain (FOG)?
Answer: Used in ANSI testing. The HFA-FOG is measured and recorded
w/ the input set at 50 dB SPL


●● Maximum Output Curve ( OSPL90)
Answer: Used in ANSI testing. It's the max output of a HA. Input is set
to 90 dB SPL to test. Notes the max at each tested frequency. HFA is
noted from this test.


●● What must be done to an aid before ANSI testing?
Answer: 1- Occlude all vents on outside of aid

, 2- Set to test mode with gain full on, output at max, and bandwidth to
max
3-Telecoil should be turned off and only turned back on when testing
SPLITS


●● What couplers do do you use with what instruments when ANSI
testing?
Answer: HA-1 = ITE
SB-1 = RIC and thin tube BTE's
HA-2 = BTE


●● What is a coupler?
Answer: A joiner for a HA that is designed off the vol of an average ear
(2cc). Used to connect the HA to a testing box for ANSI tests.


●● What is Adaptive Feedback Supression?
Answer: This allows a HA to detect and minimize feedback on it's own.


** There is also static feedback supression.


●● What is Digital Noise Reduction (DNR)?
Answer: This is an algorithm which allows a HA to seperate desired
signals ( like speech) from undesired ones ( like noise)

Información del documento

Subido en
26 de mayo de 2026
Número de páginas
157
Escrito en
2025/2026
Tipo
Examen
Contiene
Preguntas y respuestas
$13.99

¿Documento equivocado? Cámbialo gratis Dentro de los 14 días posteriores a la compra y antes de descargarlo, puedes elegir otro documento. Puedes gastar el importe de nuevo.
Escrito por estudiantes que aprobaron
Inmediatamente disponible después del pago
Leer en línea o como PDF

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
FocusFile7
3.9
(24)
Vendido
236
Seguidores
3
Artículos
62000
Última venta
7 horas hace



Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes