Medical Gas 6010 Installers Practice Test
Questions and Answers| Updated

An existing system that is not in strict compliance with the provisions of this code shall be
permitted to be continued in use as long as the _ has determined that such use does not
constitute a distinct hazard to life.

A) architect B) authority having jurisdiction C) engineer D) nursing staff

B) authority having jurisdiction




Cylinder and cryogenic liquid container outlet connections shall be designed as to render the
attachment point _ when removal of the outlet connection is attempted.

A) reusable B) unusable C) unsuitable D) universal

B) unusable




Cylinders in use and in storage shall be prevented from reaching temperatures in excess of _ *F.
A) 120 B) 125 C) 130 D) 135

B) 125




Nitrous oxide and carbon dioxide using cylinders or portable containers shall be prevented from
reaching temperatures lower than _ *F A) 130 B)125 C) 20 D) -20

C)20




Which of the following systems shall not be permitted to be located in an outdoor enclosure?

, A) bulk cryogenic liquid cylinders B) manifolds for gas cylinders C) manifolds for cryogenic liquid
containers D) medical air sources

D) medical air sources




Indoor locations for oxygen, nitrous oxide, and mixtures of these gases shall not communicate
with any of the following except _?

A) area with open flames B) designated storage area C) engines D) kitchens

B) designated storage area




Locations for central supply systems shall have _ devices protected from physical damage?

A) electrical B) pressure indicating C) regulating D) relief

A) electrical




Outdoor locations surrounded by impermeable walls, except fire barrier walls, shall have
protected ventilation openings at the _ of each wall to allow free circulation of air within the
enclosure?

A) base B) corner) C) middle D) top

A) base




All positive pressure supply systems shall be provided with duplex line pressure regulators piped
in _?

A) circuit B) loop C) parallel D) series

C) parallel

,All pressure relief valves shall be made of brass, bronze, or _?

A) carbon steel B) cast iron C) copper D) stainless steel

D) stainless steel




Central supply systems for positive pressure gases shall have one or more relief valves, the
valves shall have a relief setting that is _% above normal working operating pressure?

A) 30 B) 40 C) 50 D) 60

C) 50




When vented to the outside, relief valve vent lines shall be labeled in which way?

A) adhesive markers B) any manner that will distinguish them from the medical gas pipeline C)
stenciling D) stickers

B) any manner that will distinguish them from the medical gas pipeline




An auxiliary source connection shall consist of all of the following except?

A) removable plug or cap B) tee C) valve D) visual pressure indicator

D) visual pressure indicator




In central supply systems using gas or liquid cylinders, each header shall have a _ to prevent
debris from entering the manifold controls?

A) check valve B) filter C) monitor D) relief valve

, B) filter




Manifolds for gas cylinders shall consist of two equal headers, each with enough cylinders for an
average day's supply, but not fewer than _ connections?

A) 5 B) 4 C) 3 D) 2

D) 2




Manifolds for gas cylinders shall have a(n) _ signal that visibly indicates the operating status of
the equipment and shall activate an indicator at all master alarm panels when or just before
changeover from one header to the other takes place?

A) area B) master C) local D) zone

C) local




When multiple buildings are served from a single cryogenic liquid oxygen source, it is required
to place the emergency oxygen supply connection in _ ?

A) each building served B) the building closest to the source C) the largest building served D) the
main building only

A) each building served




The minimum clearance around EOSC for a temporary source of supply is _ ft?

A) 2 B) 3 C) 5 D) 6

B) 3

Información del documento

Subido en
21 de mayo de 2026
Número de páginas
36
Escrito en
2025/2026
Tipo
Examen
Contiene
Preguntas y respuestas
$10.99

¿Documento equivocado? Cámbialo gratis Dentro de los 14 días posteriores a la compra y antes de descargarlo, puedes elegir otro documento. Puedes gastar el importe de nuevo.
Escrito por estudiantes que aprobaron
Inmediatamente disponible después del pago
Leer en línea o como PDF

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
KelvinBrooks
4.3
(94)
Vendido
839
Seguidores
11
Artículos
5826
Última venta
1 día hace



Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes