CORRECT ANSWERS NEW UPDATE
EXAM
, receivers for medical air shall be equipped with ____ - ANSWERSrelieve valves,
automatic & manual drains, sight glass
medical air filters shall be located immediately upstream of the___ - ANSWERSfinal line
regulator
medical air compressors that are designed to separate the oil-containing section from
the compression chamber shall be provided with a coalescing filter with a(n) -
ANSWERSelement change indicator
medical air receivers shall be provided with a ____to permit servicing -
ANSWERSthree-valve bypass
dryers, filter & regulators shall be at least duplexed. Under normal operation, airflow
shall be divided equally through each device - ANSWERSFalse
medical air shall be sampled at a _____ NPS valved sample port downstream of the
final line pressure regulator - ANSWERS1/4''
Area alarm panels for medical gas systems shall indicate when line pressure in the area
being monitored increases or decreases by - ANSWERS20%
Area alarm sensors for vital life support & critical care areas shall be located on which
side of the zone valve box assemblies? - ANSWERSPatient room side, outlet inlet side,
downstream side
A signal from each local alarm shall be connected to the master alarm panels to indicate
that a problem is present with the source equipment - ANSWERStrue
A high carbon monoxide level in the medical air system shall indicate at each local
alarm site. At what level shall the alarm activate - ANSWERS10 ppm or higher
fittings for positive pressure medical gas systems must be cleaned for oxygen service
by either the manufacturers supplier or other agency prior to installation -
ANSWERStrue
the pressure in the main line of each medical gas system, the operation & condition of
the source of supply & the reserve source of supply & the reserve source (if any), shall
be monitored at the _____ alarm. - ANSWERSmaster
In a bulk cryogenic liquid system when the main supply reaches an average day's
supply, which alarm system shall be activated? - ANSWERSMaster