Verified Solutions 2025 Update
Total filtration permanently required in the useful fluoroscopic beam should not be less
than - Answers -✔✔ 2.5mm Aluminum
If X-ray source is below the table, useful beam exposure must be measured at what
distance above the table top - Answers -✔✔ 30 cm
Table top exposure rates for units without auto exposure control must be measured
every - Answers -✔✔ 3 years
Dark adaptation for the eye takes how long - Answers -✔✔ 30 minutes
Target-to-tabletop distance must be greater than and should be greater than... -
Answers -✔✔ 12 and 18 inches
Integration time for brain to recognize an object - Answers -✔✔ 0.2 seconds
Most scatter occurs at which direction from the primary beam - Answers -✔✔ 135
degrees
mA setting for spot filming where a spot film camera is not used - Answers -✔✔ 150 mA
Greatest dose to the patient comes from spot filming cameras using this film size -
Answers -✔✔ 105 mm
SI unit analogous to the rad is - Answers -✔✔ Gray
_________ is a variation in density resulting from the random spatial distribution of
quanta. - Answers -✔✔ Quantum sink
SI unit analogous to the rem - Answers -✔✔ Sievert (unit)
Exposure reduction of .25 mm lead - Answers -✔✔ 97%
Scattered radiation received is proportional to - Answers -✔✔ Dose received by the
patient
Most effective medium for static viewing in terms of reducing patient dose - Answers -
✔✔ Video disc recorder
, Contrast of the image intensifying tube is measured by - Answers -✔✔ conversion factor
maximum magnification for cinefluorography is done via which method - Answers -✔✔
total over framing
Ratio of photons that leave the output phosphor to the number of Xray photons that
strike the input phosphor - Answers -✔✔ Flux gain
Half value layer determines which characteristic of the Xray - Answers -✔✔ Xray quality
Ability to distinguish separate objects in line pairs per mm - Answers -✔✔ Resolving
power
The portion of the image intensifier responsible for magnification - Answers -✔✔
electrostatic lenses
resolution of a 9 inch image intensifier tube - Answers -✔✔ 3.5 line pairs per millimeter
light photons intercepting when electrons strike a fluorescent screen forms a... -
Answers -✔✔ Real image
Fluoro with a grid in place affects exposure rates how - Answers -✔✔ 1.5-2 x higher
Greatest radiosensitivity of fetus occurs when - Answers -✔✔ 2-8 weeks gestation
The quantity of the beam is controlled by which unit of measurement - Answers -✔✔
miliamperage
Which fluoroscopic method gives patient dose reduction of up to to 95%? - Answers -
✔✔ video disc recorder
100erg/gram is equivalent to? - Answers -✔✔ 1 rad
Fluoro units with AECs or CINE cameras must have physicist calibration how often -
Answers -✔✔ every year
Which imaging modality has the best resolution - Answers -✔✔ Radiograph
Compared to the input phosphor, the image output phosphor is... - Answers -✔✔
Smaller and brighter
As spot film size increases, what happens to the patient dose? - Answers -✔✔ Patient
dose increases